3.Instrument components and specifications
Retort |
|
Capacity: | max. 300 cassettes |
Reagent volume (in liters): | 4.3 |
Temperature (paraffin): | 40 to 65 °C |
Temperature (processing reagents): | ambient or 35 to 55 oC |
Temperature (cleaning reagents): | 50 to 65 °C |
Temperature accuracy: | + 1 K |
Filling time: | approx. 90 sec |
Drain time : | approx. 80, 120, 140 sec (adjustable) |
Impregnation vacuum: | |
Impregnation pressure: | 35 kPa (g) |
Fill vacuum: | |
Drain pressure: | 35 kPa (g) |
|
|
General |
|
Reagent bottles | 10 |
Cleaning solution bottles: | 3 |
Maximum bottle volume: | 4.3 l |
Printer: | Optional accessory |
Pretest check: | ON/OFF |
Fluid level sensor: | ON/OFF |
Recirculation (pump in/out): | ON/OFF |
(a) Time before 1st cycle: | 12 min |
(b) Time between cycles: | 20 min |
|
|
System setup |
|
Password status: | Supervisor / User |
Type of password: | alphanumeric, freely selectable |
Reagent Management Status: | ON/OFF |
Software access control (‘Lock Mode’): | ON/OFF |
16 | Instruction manual V 1.3 – 12/2003 |