Leica ASP300 instruction manual Adding molten paraffin, Operation

Page 33
Adding molten paraffin

5. Operation

Adding molten paraffin

Molten wax may have a maximum temperature of 70 °C.

If filling with molten wax, do not fill above the MAX level (see Fig. 15).

After a wax bath has been filled, it must be marked as ‘full’ in the REAGENT STATUS screen.

Software control

If the wax baths are filled with molten wax, it is possible to override the Leica ASP300 melting time calculation.

For that purpose, in supervisor mode, access the

SERVICE FUNCTIONS screen (Fig. 16).

 

 

• Press TIME UNTIL WAX MELTED.

• Enter the required delay time and press OK to confirm.

Fig. 16

If the instrument attempts a paraffin step with the paraffin only partly molten, severe instru- ment malfunctions may occur.

Leica ASP300

33

Image 33
Contents Automated vacuum tissue processor Leica ASP300Advanced Smart Processor Page Published by Table of contents Trouble shooting Table of contentsCleaning and maintenance 7.3.1 Daily cleaning and maintenance1.Important notes 1.1Symbols used in this manual and their meaning1.2 Designated use Chapter 3.5 - “Compatible reagents“2.1 Safety instructions 2. Safety2.2 Warnings 2.Safety2. Safety Warnings - cleaning and maintenanceWarnings - working with the instrument Overpressure protection 2.3 Integrated safety devicesProtection against excessive vacuum Overcurrent protection3.1Specific instrument options 3.Instrument components and specifications3.Instrument components and specifications 3.2 Overview - instrument componentsInstrument components and accessories Overview - Instrument components3.Instrument components and specifications 3.Instrument components and specifications 3.3 Standard delivery - packing list3.4 Technical data Instrument components and specificationsRetort 3.Instrument components and specifications3.Instrument components and specifications 3.4 Technical Data continued3.Instrument components and specifications 3.5 Compatible reagents4.2Electrical connection 4. Setting up the instrument4.1 Site requirements Moving the instrument4. Setting up the instrument Connecting the instrument to mains4.3 Installing the accessories 4.Setting up the instrument4.Setting up the instrument 4.4 Setting up the data connectionsPrinter optional Lift off the instrument console flap4. Setting up the instrument 4.5 Alarm functions4.Setting up the instrument 4.6 Switching the instrument on2 - Access INSTALLATION menu 4.Setting up the instrumentButton symbols 4.Setting up the instrument4.Setting up the instrument 4.7.2. Access control→→ →→ →→ →→ → → → → →→→ →→4.Setting up the instrument 5. Operation 5.1 Reagent handling5.1.1Fill/drain reagents other than wax Fill / drain reagents → 5.Operation5.1.2 Changing the wax → + →5.Operation 5.Operation Filling with wax pellets → →5. Operation Adding molten paraffin5.2 Setting up the instrument parameters → 5.2.1 System setup5.Operation Program Options 5. OperationSecurity 5.Operation5. Operation 5.2.2 Editing the reagent listEntering / modifying reagent thresholds 5. Operation5.Operation ↓ ↓ ↓5.Operation 5.2.3Viewing the program list5. Operation 5.2.4 Adding and/or modifying programs5.Operation Editing program steps continued 5. Operation5.Operation 5.2.5Favorites maximum of 5 programs5.Operation 5.Operation 5.2.6 Editing the stations5. Operation 5.Operation 5.3 ProgramsEDIT to access the Edit Program screen 5.3.1 Editing a selected programEnter the day when you want the program finished 5. Operation5.Operation LEVEL CHECK enabled5. Operation 5.3.2 Starting a program5.Operation 5.4Reagent status →5. Operation 5.5 System Monitor → →5.Operation 5.6 On-linehelp5.Operation 5.7Configuring a printer optional →→6.Trouble shooting → 6.2 Power failure6.1 General 6.3Trouble shooting → 6.Trouble shooting6.Trouble shooting 6.4 Typical fill or drain problems7.1Clean programs →→ 7.1.1 Retort clean programs 7.Cleaning and maintenance7.Cleaning and maintenance Editing steps7.1.2 Wax clean programs 7. Cleaning and maintenance7.Cleaning and maintenance 7.1.3Smart clean →7. Cleaning and maintenance 7.2 General cleaning stepsDraining the condensate bottle 7.Cleaning and maintenanceExchanging the activated carbon filter Exchange the filter as shown in Fig7. Cleaning and maintenance 7.3.1Daily cleaning and maintenance7.Cleaning and maintenance 7.3.2Periodic cleaning and maintenance7. Cleaning and maintenance 7.4 Preventative maintenance scheduleMaintenance activity 7. Cleaning and maintenanceWarranty 8. Warranty and serviceTechnical service information Decommissioning and disposal