Leica Table of contents, Leica ASP300, Trouble shooting, Cleaning and maintenance

Page 5
Trouble shooting

 

 

 

Table of contents

 

 

 

 

 

5.2.6

Editing the stations

46

 

5.2.7

Reagent groups

47

 

5.3

Programs

48

 

5.3.1 Editing a selected program

49

 

5.3.2

Starting a program

51

 

5.4

Reagent status

52

 

5.5

System Monitor

53

 

5.6

On-line help

54

 

5.7

Configuring a printer (optional)

55

6.

Trouble shooting

56

 

6.1

General

56

 

6.2

Power failure

56

 

6.3

Trouble shooting

57

 

6.4

Typical fill or drain problems

58

7.

Cleaning and maintenance

59

 

7.1

Clean programs

59

 

7.1.1

Retort clean programs

59

 

7.1.2

Wax clean programs

61

 

7.1.3

Smart clean

62

 

7.2

General cleaning steps

63

 

7.3.1 Daily cleaning and maintenance

65

 

7.3.2 Periodic cleaning and maintenance

66

 

7.4

Preventative maintenance schedule

67

8.

Warranty and service

69

Leica ASP300

5

Image 5
Contents Advanced Smart Processor Automated vacuum tissue processorLeica ASP300 Page Published by Table of contents Trouble shooting Table of contentsCleaning and maintenance 7.3.1 Daily cleaning and maintenance1.Important notes 1.1Symbols used in this manual and their meaning1.2 Designated use Chapter 3.5 - “Compatible reagents“2.1 Safety instructions 2. Safety2.2 Warnings 2.SafetyWarnings - working with the instrument 2. SafetyWarnings - cleaning and maintenance Overpressure protection 2.3 Integrated safety devicesProtection against excessive vacuum Overcurrent protection3.1Specific instrument options 3.Instrument components and specifications3.Instrument components and specifications 3.2 Overview - instrument components3.Instrument components and specifications Instrument components and accessoriesOverview - Instrument components 3.Instrument components and specifications 3.3 Standard delivery - packing list3.4 Technical data Instrument components and specificationsRetort 3.Instrument components and specifications3.Instrument components and specifications 3.4 Technical Data continued3.Instrument components and specifications 3.5 Compatible reagents4.2Electrical connection 4. Setting up the instrument4.1 Site requirements Moving the instrument4. Setting up the instrument Connecting the instrument to mains4.3 Installing the accessories 4.Setting up the instrument4.Setting up the instrument 4.4 Setting up the data connectionsPrinter optional Lift off the instrument console flap4. Setting up the instrument 4.5 Alarm functions4.Setting up the instrument 4.6 Switching the instrument on2 - Access INSTALLATION menu 4.Setting up the instrumentButton symbols 4.Setting up the instrument4.Setting up the instrument 4.7.2. Access control4.Setting up the instrument →→ →→ →→ →→ → → → →→→→ →→ 5.1.1Fill/drain reagents other than wax 5. Operation5.1 Reagent handling Fill / drain reagents → 5.Operation5.Operation 5.1.2 Changing the wax→ + → 5.Operation Filling with wax pellets → →5. Operation Adding molten paraffin5.Operation 5.2 Setting up the instrument parameters→ 5.2.1 System setup Program Options 5. OperationSecurity 5.Operation5. Operation 5.2.2 Editing the reagent listEntering / modifying reagent thresholds 5. Operation5.Operation ↓ ↓ ↓5.Operation 5.2.3Viewing the program list5. Operation 5.2.4 Adding and/or modifying programs5.Operation Editing program steps continued 5. Operation5.Operation 5.2.5Favorites maximum of 5 programs5.Operation 5.Operation 5.2.6 Editing the stations5. Operation 5.Operation 5.3 ProgramsEDIT to access the Edit Program screen 5.3.1 Editing a selected programEnter the day when you want the program finished 5. Operation5.Operation LEVEL CHECK enabled5. Operation 5.3.2 Starting a program5.Operation 5.4Reagent status →5. Operation 5.5 System Monitor → →5.Operation 5.6 On-linehelp5.Operation 5.7Configuring a printer optional →→6.1 General 6.Trouble shooting→ 6.2 Power failure 6.3Trouble shooting → 6.Trouble shooting6.Trouble shooting 6.4 Typical fill or drain problems7.1Clean programs →→ 7.1.1 Retort clean programs 7.Cleaning and maintenance7.Cleaning and maintenance Editing steps7.1.2 Wax clean programs 7. Cleaning and maintenance7.Cleaning and maintenance 7.1.3Smart clean →7. Cleaning and maintenance 7.2 General cleaning stepsDraining the condensate bottle 7.Cleaning and maintenanceExchanging the activated carbon filter Exchange the filter as shown in Fig7. Cleaning and maintenance 7.3.1Daily cleaning and maintenance7.Cleaning and maintenance 7.3.2Periodic cleaning and maintenance7. Cleaning and maintenance 7.4 Preventative maintenance scheduleMaintenance activity 7. Cleaning and maintenanceWarranty 8. Warranty and serviceTechnical service information Decommissioning and disposal