5.Operation
↓
↓
↓
Deleting reagents
•Highlight the reagent to be deleted in the SET UP REAGENTS AND WARNING THRESHOLDS screen.
•Press DELETE.
•Press OK in the screen prompt.
Please remember that a reagent which is already used in a pro- gram cannot be deleted.
Adding new reagents to stations
Go to SMART FUNCTIONS to
If a bottle is filled manually, it must be defined as full afterwards. For that purpose:
•Press MENU to access the MENU FUNCTIONS screen.
•Press REAGENT STATUS.
•In the reagent list, highlight the station / reagent that has been filled manually.
•Press SET AS FULL to mark the station as ‘full’.
Upon activation of the RMS all warning thresholds for the se- lected reagent are automatically reset to ‘0’.
Leica ASP300 | 39 |