Teledyne HVC-282 instruction manual 5.3Check Against Primary Standard

Page 11

f)Adjust outside pot to read 2.630mV ±5V on DVM.

g)DV-273 gauge tube is now calibrated.

HVC-282calibration.

Courseadjustment

(Caution:Resistorconnectionsarenottoanypowerterminals.)

a)Withpoweroff,connectan86.6ohm1%1/8Wresistortotubeterminalswhite&green.

b)Connecta28ohm1%1/8Wresistortotubeterminalsredandblack.

c)ConnectaDVMtomeasuremVacrossthe28ohmresistorwith(+)leadtoredterminal.

d)Withpoweron,adjusttrimpotonrightsideofcan(Viewedfromfrontdisplay.)

to set 350mV ±0.5mV.

e)Setunitsdisplaytoread0.00byadjustingzeropotatrearofcan.

f)Withpoweroff,attacha2.630mV±5VDCsourceacross(+)white(-)greenthe86.6 ohm resistor.

g)Withpoweronadjustspanpotatrearofcantilldisplayreads15.00Torr.

Hastings) HVC-282 is now coarsely calibrated.

Fine adjustment.

Due to bridge loading, you will need to readjust zero and span pots using a good or above calibrated DV- 273.

a)Connect and run unit for 30 minutes at hard vacuum (< 1mTorr) so system may out gas.

b)Readjust zero pot at rear of can if necessary, to set display to read 0.00 Torr.

c)Using a Primary standard, bring system up to 15.000 Torr ±1mTorr.

d)Readjust span pot at rear of can if necessary, to set display to read 15.00 Torr.

e)Fine Calibration is now complete.

5.3Check Against Primary Standard:

If additional calibrations of the vacuum gauge are desired, an accurate measuring device such as the McLeod Gauge, spinning rotor gauge, or capacitance manometer may be employed as a reference. Please refer to manufacture for the suitability requirements. Please read Sec. 6.

PAGE 11

Image 11
Contents MANUAL INSTRUCTIONMODEL HVC-282 VACUUM CONTROLLER Manual Print History TABLE OF CONTENTS PAGE 1.2 OperatingPrinciple 1.3 Specifications1.0 GENERAL INFORMATION 1.1 Description1.4 Accessories FIG.2.1 2.0GAUGE TUBE AND CABLES2.1GaugeTubes 2.2 GaugeTubeCables3.1PanelMounting 3.0INSTALLATION3.2 GaugeTubeInstallation 4.0 OPERATION PAGE 5.0 CALIBRATION AND TROUBLESHOOTING GUIDE5.1 CheckingofTube 5.2 Calibration5.3Check Against Primary Standard 6.3 Effect of Thermal Conductivity 6.0NOTES ON VACUUM MEASUREMENT6.1Effects of Condensable Vapors 6.2 Outgassing7.2 Non-WarrantyRepair Policy 7.0 SERVICE INFORMATION7.1 Warranty Repair Policy PAGE 7.3ServiceForm PAGE PAGE 8.0 DRAWINGS and DIAGRAMSPAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE PAGE