sequence the status LED alternates flashes of RED and GREEN, the analog output changes back and forth from 3.0 to 3.3mA and Fail Cal. message is displayed on the screen. The unit remains in this state until acknowledged by a manual Reset. After manual Reset the program will return to the normal operation mode using previous calibration values. Since the calibration was unsuccessful another attempt may be made or replace the sensor and re-try calibration.

Calibration Procedure for UNI-TRAN Premium

Plus ST1400 O2 sensor

Use clean ambient air as calibration gas otherwise, use certified instrument air.

Be sure the UNI-TRAN Premium Plus is powered-up and is not indicating a fault; FAULT/CAL LED is showing a short green flash every 2.0 seconds (Confidence blip).

Ensure that the sensor is in a clean air environment before beginning the calibration procedure.

To enter the main menu, activate the magnetic reed switch or manual push button switch. When the countdown from 5 to 0 is complete remove the magnet from the magnetic reed switch or release the CAL/RESET push button. When prompted with Calibrate sensor “YES?” select the function by momentarily placing the magnet on the reed switch or depressing the CAL/RESET button. The selection is acknowledged with flashing “YES”. The controller starts calibration. The following Instructions are scrolled across the display for calibration:

Apply Calibration Gas - at this time the input should be ambient air (3.0mA output current level).

Processing - The sensor is reading the ambient air level (3.0mA output current level).

20.9- The sensor is calculating the corrected oxygen level (3.6mA output current level).

Remove Calibration Air - at this time input should be reduced to 3.6 mA by removing span gas from the sensor if it is attached (3.6mA output current level).

Calibration Complete - Returns to normal operation (17.38 mA output current)

NOTE:

If span setting does not complete successfully within ten minutes of starting the calibration sequence the status LED alternates flashes of RED and GREEN, the analog output changes back and forth from 3.0 to 3.3mA and Fail Cal. message is displayed on the screen. The unit remains in this state until acknowledged by a manual Reset. After manual Reset the program will return to the normal operation mode using previous calibration values. Since the calibration was unsuccessful another attempt may be made or replace the sensor and re-try calibration.

22

Image 28
Contents UNI-TRAN Premium Plus Toxic GAS Detector UT-P+-STXXXX Page Page Page Table of Contents Page Introduction FeaturesIntroduction Installation and Start Up What’s in the packageLocation of Sensors UnpackingMounting Toxic Detector LayoutDimensional Drawings Page Page Page Wiring Sensor Separation Page Wiring Diagrams Page Page Page Page Page Installation Checklist Start UpRelay Settings Default Relay SettingsSummary of Main Menu Exit Return to normal if no option is selectedReview Settings System CalibrationCalibration Procedure Page Periodic Response Check OperationOperator Interface Calibration Procedure Table of ResponsesMagnetic Reed Switch Activation Relay Settings Review Settings Start Delay Sensor FaultHardware Sensor DriftRelays 20 mATroubleshooting Sensor LifeGlossary of Terms Ordering Information Published in a ccordance w ith E1A standard Appendix B Wire Resistance In Ohms AWGAppendix C Specifications Sensor SpecificationsPage