Wiring the Transmitter to the Sensor

3.1Overview

This chapter describes how to connect Micro Motion Model 1700 and 2700 transmitters to a Micro Motion sensor.

Note: If you have an integral installation, this step is not required. Continue with wiring the transmitter outputs (Chapters 4–7).

Wiring requirements between the sensor and transmitter depend on the installation type (see Figure 2-1).

If you have a 4-wire remote transmitter installation, review the information on 4-wire cable in Section 3.2, then follow the instructions in Section 3.3.

If you have a 9-wire remote transmitter installation, review the information on 9-wire cable in Section 3.2, then follow the instructions in Section 3.4.

If you have a remote core processor with remote transmitter installation, review the information on both 4-wire and 9-wire cable in Section 3.2, then follow the instructions in Section 3.5.

CAUTION

Large electromagnetic fields can interfere with flowmeter communication signals.

Improper installation of cable or conduit can cause measurement error or flowmeter failure. To reduce the risk of measurement error or flowmeter failure, keep cable or conduit away from devices such as transformers, motors, and power lines which produce large electromagnetic fields.

3.2Cable types

This section describes the types of 4-wire cable and 9-wire cable that can be used for wiring the transmitter to the sensor.

Before You Begin

Installing the Transmitter

Sensor Wiring

Output Wiring – Analog

Installation Manual

19

Image 23
Contents Micro Motion Model 1700 Model 2700 Transmitters Page Contents Index Overview SafetyBefore You Begin Flowmeter components Flowmeter documentation 1Flowmeter documentation resources Transmitter installation proceduresMicro Motion customer service Installation architecture Installing the Transmitter1Installation types Determining an appropriate location Temperature limitsPower source AC power requirements1Typical power cable resistances at 68 F 20 C DC power requirementsMounting the transmitter Accessibility for maintenanceMaximum cable lengths 2Maximum cable lengths2Rotating the transmitter Integral installations34-wire remote Wall mount or pipe mount 59-wire remote Wall mount or pipe mount 3 9-wire remote installationsMounting the remote core processor 6Transmitter/core processor assembly Exploded view7Remote core processor Wall mount or pipe mount 3Grounding methods for flowmeter components Grounding the flowmeter componentsRotating the display Supplying powerInstalling the Transmitter 10Display components Cable types Wiring the Transmitter to the SensorWiring for 4-wire remote installations 1 4-wire cable2 9-wire cable 3Wiring to the mating connector Core processor on transmitter Wiring for 9-wire remote installationsSubtask 1 Wire the remote core processor to the transmitter 5Cable glands 7Wrapping the shield drain wires 10Gland body and core processor housing 129-wire cable between sensor and core processor Subtask 2 Wiring the sensor to the remote core processorWiring the Transmitter to the Sensor 1Terminals and output types Output Wiring Model 1700/2700 Analog TransmittersOutput wiring Output terminals and output types1Basic analog wiring 2HART/analog single-loop wiring Smart Family Output Wiring Intrinsically Safe Safe area mA output wiring Safe area output wiring2Safe area mA output load resistance values 5Safe area frequency/discrete output wiring Safe area frequency/discrete output wiringHazardous area safety parameters Hazardous area output wiringHazardous area capacitance 2Safety parametersHazardous area voltage Hazardous area currentHazardous area frequency/discrete output wiring Hazardous area mA output wiringHazardous area Channel configuration Output Wiring Model 2700 Configurable I/O TransmittersMA output wiring 1Channel configuration2Basic mA wiring Frequency output wiring 5Frequency output Terminals 3 & 4 Channel B Internal power 7Frequency output Terminals 5 & 6 Channel C Internal power 9Discrete output 1 Terminals 3 & 4 Channel B Internal power Discrete output wiring11Discrete output 2 Terminals 5 & 6 Channel C Internal power Volts Voltage Output Level High 2Input voltage ranges for external power Discrete input wiring16Discrete input Terminals 5 & 6 Channel C Internal power Micro Motion Model 1700 and 2700 Transmitters 1Connecting the fieldbus communication wires Foundation fieldbus wiring2Connecting the PROFIBUS-PA communication wires PROFIBUS-PA wiringService port connection SpecificationsFunctional specifications Power connectionFisco and Fnico approval Input/output signals Input signal from sensorCore processor connection Specifications Operating Absolute minimum = 100 Ohms for V supply 25.6 volts Specifications Foundation fieldbus or PROFIBUS-PA transmitters Service portDigital communication HART/Bell202EMC compliance Power supplyEnvironmental requirements Ambient temperature limits Ambient temperature effectAtex and IECEx UL and CSATransmitter OutputsMounting Performance specificationsPhysical specifications HousingInterface/display WeightDimensions Dimensions in inchesmm Inches Dimensions in mm Or M20 × 11/16 160 13/16 139 ″-14 NPT 158 182 × Ø3/8 13/16 148 15/16 × 1/2″-14 NPT Figure A-5Dimensions Remote core processor Atex IndexRemote core processor with remote Index Micro Motion Model 1700 and 2700 Transmitters Page 20001700