Specifications
A.2 Hazardous area classifications
The transmitter may have a tag listing hazardous area classifications, which indicate suitability for installation in the hazardous areas described in this section.
A.2.1 UL and CSA
Ambient temperature is limited to
Ambient temperature is limited to
Transmitter
Class I, Division 1, Groups C and D. Class II, Division 1, Groups E, F, and G explosion proof (when installed with approved conduit seals). Otherwise, Class I, Division 2, Groups A, B, C, and D.
Outputs
Provides nonincendive sensor outputs for use in Class I, Division 2, Groups A, B, C, and D; or intrinsically safe sensor outputs for use in Class I, Division 1, Groups C and D or Class II, Division 1, Groups E, F, and G.
A.2.2 ATEX and IECEx
Ambient temperature is limited to below 131 °F (55 °C) for ATEX and IECEx compliance. Table
Table A-1 ATEX and IECEx classifications
Output option | Classification | ATEX | IECEx | |
|
|
|
| |
Analog, | Flameproof when installed with |
|
| |
approved cable glands: |
|
| ||
FOUNDATION fieldbus, | • With display | EEx d [ib] IIB+H2 T5 | Ex d [ib] IIB+H2 T5 | |
and configurable I/O | ||||
|
|
| ||
(output options A, N, B, and C) | • Without display | EEx d [ib] IIC T5 | Ex d [ib] IIC T5 | |
|
|
|
| |
| Increased safety when installed |
|
| |
| with approved cable glands: |
|
| |
| • With display | EEx de [ib] IIB+H2 T5 | Ex de [ib] IIB+H2 T5 | |
| • Without display | EEx de [ib] IIC T5 | Ex de [ib] IIC T5 | |
|
|
|
| |
Intrinsically safe analog, | Flameproof when installed with |
|
| |
intrinsically safe FOUNDATION | approved cable glands: |
|
| |
fieldbus, and | • With display | EEx d [ia/ib] IIB+H2 T5 | Ex d [ia/ib] IIB+H2 T5 | |
(output options D, E, and G) | ||||
|
|
| ||
| • Without display | EEx d [ia/ib] IIC T5 | Ex d [ia/ib] IIC T5 | |
|
|
|
| |
| Increased safety when installed |
|
| |
| with approved cable glands: |
|
| |
| • With display | EEx de [ia/ib] IIB+H2 T5 | Ex de [ia/ib] IIB+H2 T5 | |
| • Without display | EEx de [ia/ib] IIC T5 | Ex de [ia/ib] IIC T5 | |
|
|
|
|
Output Wiring – Intrinsically Safe
Output Wiring – Configurable I/O Output Wiring – Fieldbus/PROFIBUS
Specifications
Installation Manual | 63 |