File Management

If the log rate is greater than or equal to 5 seconds, OMWT-SOFT-OPC opens and then closes the log file each time data is logged. If the log rate is less than 5 seconds, OMWT-SOFT-OPC leaves the file open and closes the file when OMWT-SOFT- OPC is terminated, the log rate is set to zero, or the log rate is set to 5 or greater. The log file can be viewed while OMWT-SOFT-OPC is logging. The other program must not write to the file, lock the file, set the file as read only, or set the file's sharing parameters so that OMWT-SOFT-OPC cannot log.

When processing data with another program, the best approach is to either rename the file if the log rate is 5 or greater (OMWT-SOFT-OPC will automatically create a new file on the next log) or make a copy of the log file. When renaming, you may get an access error. Just try again.

The Error log file is opened and then closed whenever OMWT-SOFT-OPC writes an error record.

48 N OMWT-SOFT-OPC Server Data Logging

OMWT-SOFT-OPC Server

Image 52
Contents OWMT-SOFT-OPC OMEGAnet Online Service Internet e-mail Info@omega.com Contents OMWT-SOFT-OPC Server Data Logging Overview Quick StartOverview OMWT-SOFT-OPC Server Wireless Setup OMWT-SOFT-OPC Server and WirelessDescription Edit Device Receivers Device ListOMWT-RECIP OMWT-REPEATPolling Identification FilteringWhat is OPC? Data Logging DiagnosticsSetup OMWT-SOFT-OPC Main WindowMain Window Change PortAdd/Delete IP Connection View IP Connection Status Outputs New and EditTrain ReConfigure Net Search and Add DevicesQuit Devices ErrorsHelp OMWT-SOFT-OPC Server .INI File Format OMWT-SOFT-OPC.iniServer Logging IP AddressesTemperature Setup Humidity Digcounter and Directional Counter DirectcntCnttemp Fastcnttemp FASTCNTTEMP1Alarmtemp OMWT-AIN5V, OMWT-AIN10V, OMWT-AIN20MA, OMWT-VIB, Omwt Analog InputMOT Access/Control Reader Repeater Dual Discrete OutputWireless Analog Output OUTPUTANALOG1OMWT-SOFT-OPC INI Backup OMWT-SOFT-OPC Server Tags Device NamesOMWT-SOFT-OPC Server Tags Temperature Device Tag NamesDigital I/O DIGITALn Temperature/Humidity HUMIDITYn OMWT-IRCNT, OWT-DIRCNT Cnttemp CNTTEMPn Fastcnttemp FASTCNTTEMPn Alarmtemp ALARMTEMPn AnalogInput AIn Access/Control Reader IDRn OMWT-SOFT-OPC Server Tags Repeater REPEATERn Dual Discrete Output OUTPUTDISCn Wireless Analog Output OutputAnalogn Server Item Names OMWT-SOFT-OPC Server OMWT-SOFT-OPC Server Tags N OMWT-SOFT-OPC Server Tags Features OMWT-SOFT-OPC Server Data LoggingFile Management File Format Data Log File Format Error Log OMWT-RECIP Receivers Setting up to use OMWT-RECIP ReceiversOverview of OMWT-RECIP Receivers OMWT-RECIP Receiver Notes Error Messages Communication ErrorsCommunication Error CRC16 or Time Out Runtime ErrorsOMWT-SOFT-OPC Server Internal Error Why Use Corrections CorrectionsCorrection IDs Pre-defined ThermistorHow to use Corrections Correction.INI filePredefined Corrections NtcnormalresWARRANTY/DISCLAIMER OMEGA…Of Course