Omega Engineering M-4066 manual Predefined Corrections, Ntcnormalres

Page 61

Predefined Corrections

The device type "Thermistor" will default to a correction ID of 128. To override this default you can add a CorrID parameter to the device section of the OMWT-SOFT- OPC.INI file, or you can set a new default for all thermistors in the Server section of the OMWT-SOFT-OPC.INI file. To change the default CorrectionID for all thermistors add the parameter "DefaultThermistorCorrID=" to the Server section, and set it to whatever CorrectionID you wish.

Predefined Corrections use the following equations. The first equation normalizes the passed value, and the second produces the conversion to degrees Celsius. Parameters:

NTC_DIVIDERES = 10000

NTC_NORMALRES = 10000

Fitting constants:

A = 3.3539264E-3

B = 2.5609446E-4

C = 1.9621987E-6

D = 4.6045930E-8

Rnormal = NTC_DIVIDERES / (32767

/ value ) - 1

/

NTC_NORMALRES

Value in degrees Celsuis = 1 / ( A

+ B*Rnormal

+

C*Rnormal2 +

D*Rnormal3

)

 

 

 

OMWT-SOFT-OPC Server

Corrections N 57

Image 61
Contents OWMT-SOFT-OPC OMEGAnet Online Service Internet e-mail Info@omega.com Contents OMWT-SOFT-OPC Server Data Logging Quick Start OverviewOverview OMWT-SOFT-OPC Server Wireless Setup OMWT-SOFT-OPC Server and WirelessDescription Edit Device Device List ReceiversOMWT-RECIP OMWT-REPEATIdentification Filtering PollingDiagnostics What is OPC? Data LoggingOMWT-SOFT-OPC Main Window SetupMain Window Change PortAdd/Delete IP Connection View IP Connection Status New and Edit OutputsTrain Search and Add Devices ReConfigure NetQuit Devices ErrorsHelp OMWT-SOFT-OPC.ini OMWT-SOFT-OPC Server .INI File FormatServer IP Addresses LoggingTemperature Setup Humidity Directcnt Digcounter and Directional CounterCnttemp FASTCNTTEMP1 FastcnttempAlarmtemp OMWT-AIN5V, OMWT-AIN10V, OMWT-AIN20MA, OMWT-VIB, Omwt Analog InputMOT Access/Control Reader Dual Discrete Output RepeaterOUTPUTANALOG1 Wireless Analog OutputOMWT-SOFT-OPC INI Backup Device Names OMWT-SOFT-OPC Server TagsOMWT-SOFT-OPC Server Tags Device Tag Names TemperatureDigital I/O DIGITALn Temperature/Humidity HUMIDITYn OMWT-IRCNT, OWT-DIRCNT Cnttemp CNTTEMPn Fastcnttemp FASTCNTTEMPn Alarmtemp ALARMTEMPn AnalogInput AIn Access/Control Reader IDRn OMWT-SOFT-OPC Server Tags Repeater REPEATERn Dual Discrete Output OUTPUTDISCn Wireless Analog Output OutputAnalogn Server Item Names OMWT-SOFT-OPC Server OMWT-SOFT-OPC Server Tags N OMWT-SOFT-OPC Server Tags OMWT-SOFT-OPC Server Data Logging FeaturesFile Management File Format Data Log File Format Error Log OMWT-RECIP Receivers Setting up to use OMWT-RECIP ReceiversOverview of OMWT-RECIP Receivers OMWT-RECIP Receiver Notes Communication Errors Error Messages Communication Error CRC16 or Time Out Runtime Errors OMWT-SOFT-OPC Server Internal Error Corrections Why Use CorrectionsCorrection IDs Pre-defined ThermistorCorrection.INI file How to use CorrectionsNtcnormalres Predefined CorrectionsWARRANTY/DISCLAIMER OMEGA…Of Course