START-UP OF UNIT
Follow the procedures for the initial and consequent operations of the unit.
1.Connect the power cord to the dedicated outlet with appropriate rating.
2.On
buzzer stop key (BUZZER). The alarm will operate until the temperature monitor bottle sensor goes into the range of 4oC±2oC.
3.Ensure that both monitor bottles have been filled with 10% glycerol or water as outlined on page 13.
4.The unit has been factory preset to operate at 4oC±1oC. There is no need to set the temperature.
5.It takes several hours to stabilize the chamber temperature at the set temperature. Once the unit has stabilized, ensure the monitor bottle temperature corresponds with the set temperature.
6.Turn on the fluorescent light switch to check the light. After checking, turn off the switch if the lighting is not necessary.
7.Once the unit operation has been thoroughly checked, begin slowly placing items into the chamber to minimize the temperature rise.
Operation after power failure
The refrigerator resumes the operation with temperature setting of 4oC after power failure.
15