Sanyo MBR-304DR instruction manual START-UP of Unit, Operation after power failure

Page 16

START-UP OF UNIT

Follow the procedures for the initial and consequent operations of the unit.

1.Connect the power cord to the dedicated outlet with appropriate rating.

2.On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the

buzzer stop key (BUZZER). The alarm will operate until the temperature monitor bottle sensor goes into the range of 4oC±2oC.

3.Ensure that both monitor bottles have been filled with 10% glycerol or water as outlined on page 13.

4.The unit has been factory preset to operate at 4oC±1oC. There is no need to set the temperature.

5.It takes several hours to stabilize the chamber temperature at the set temperature. Once the unit has stabilized, ensure the monitor bottle temperature corresponds with the set temperature.

6.Turn on the fluorescent light switch to check the light. After checking, turn off the switch if the lighting is not necessary.

7.Once the unit operation has been thoroughly checked, begin slowly placing items into the chamber to minimize the temperature rise.

Operation after power failure

The refrigerator resumes the operation with temperature setting of 4oC after power failure.

15

Image 16
Contents MBR-304DR MBR-304GR MBR-304D MBR-304G Contents Precautions for Safe Operation Symbol showsPrecautions for Safe Operation Prevent electric shock or injury Damage Page Environmental Conditions Refrigerator Components Refrigerator Components Control panel and keypad Installation Installation siteInstallation Preparation of monitor bottle Before Commencing OperationLoosen Remote alarm terminal Operation after power failure START-UP of UnitOperating Instructions Storing of itemsAir-intake vent Air-exhaust ventTemperature display Defrost cyclesCycle defrost Forced defrostCalibration of temperature sensor Alarms and safety functions Alarms & Safety FunctionsAlarm test procedure Cleaning of cabinet Routine MaintenanceReplacement of fluorescent lamp Cleaning of condenser filter Replacement of glow starterDisposal of Unit Troubleshooting If nothing operates even when switched onAlarm device is activated When the unit does not get cold enoughSpecifications PerformanceSafety check sheet CEM