Sanyo MBR-304DR Temperature display, Defrost cycles, Cycle defrost, Forced defrost

Page 18

OPERATING INSTRUCTIONS

Temperature display

The unit is set at the factory to obtain the proper temperature automatically (4oC±1oC). There is no need to set the temperature.

Table 1. Temperature display

: Lamp ON : Lamp OFF

 

 

Operation

Key operated

Display

 

Display mode

 

Plug-in the power

 

 

 

 

Average monitor bottle

UPPER

 

1

supply cord and turn on

-----

 

 

Average temperature

 

 

temperature

LOWER

 

the power switch.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press the temperature

 

 

 

 

Upper monitor bottle

UPPER

 

2

 

UPPER

Upper temperature

display key.

 

 

 

 

temperature

LOWER

 

LOWER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press the temperature

 

 

 

Lower monitor bottle

UPPER

 

3

 

UPPER

Lower temperature

display key.

 

 

 

 

temperature

LOWER

 

 

LOWER

 

 

 

 

 

 

 

 

 

 

 

 

Press the temperature

 

 

 

Average monitor bottle

UPPER

 

4

 

UPPER

Average temperature

display key.

 

 

 

 

temperature

LOWER

 

 

LOWER

 

 

 

 

 

 

 

 

 

 

 

5

Repeat from step 2.

 

 

 

 

 

 

 

Note:

The displayed temperature represents the temperature sensed by a thermal sensor attached to the monitor bottle mounted to the upper and lower portion of the chamber.

The displayed temperature may not always be at 4oC. Because the chamber probes are representative of high and low average temperatures within the chamber.

Defrost cycles

There is no need for routine defrosting of the unit as this occurs automatically as follows:

1. Cycle defrost

To keep the chamber temperature stable, the refrigeration compressor is cycled on and off. During offperiod any frost which has accumulated on the evaporator is melted by energizing a defrost heater. This will not have any discernible effect on the chamber temperature.

2. Forced defrost

When the ambient humidity is high, or a large amount of damp product is being stored inside the chamber, there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In this case, a forced defrost cycle can be initiated.

When the unit is operating under a forced defrost cycle, the current chamber temperature and dFis displayed alternately on the digital temperature display.

Once the forced defrost cycle is completed, normal operation resumes.

The temperature of blood bag (400mL) in the chamber rises by approximately 1 to 2oC during the defrosting.

17

Image 18
Contents MBR-304DR MBR-304GR MBR-304D MBR-304G Contents Precautions for Safe Operation Symbol showsPrecautions for Safe Operation Prevent electric shock or injury Damage Page Environmental Conditions Refrigerator Components Refrigerator Components Control panel and keypad Installation Installation siteInstallation Before Commencing Operation Preparation of monitor bottleLoosen Remote alarm terminal Operation after power failure START-UP of UnitAir-intake vent Operating InstructionsStoring of items Air-exhaust ventCycle defrost Temperature displayDefrost cycles Forced defrostCalibration of temperature sensor Alarms & Safety Functions Alarms and safety functionsAlarm test procedure Routine Maintenance Cleaning of cabinetReplacement of fluorescent lamp Replacement of glow starter Cleaning of condenser filterDisposal of Unit Alarm device is activated TroubleshootingIf nothing operates even when switched on When the unit does not get cold enoughSpecifications PerformanceSafety check sheet CEM