BEFORE COMMENCING OPERATION

Preparation of monitor bottle

Prior to the operation of the unit, fill the monitor bottles (upper and lower) with 10% glycerol or water using the following procedure:

1.Remove the top and bottom drawer (or shelf).

2.Remove the monitor bottle cover as shown in the figure 1.

3.Take out the sensor from the monitor bottle.

4.Remove the fixing screw and take out the monitor bottle.

5.Remove the cap by loosing it as shown in the figure 2.

6.Pour 200ml of 10% glycerol or water into the monitor bottle.

7.Cap the monitor bottle and replace the bottle to its original position. Then set the sensor to the bottle and replace the cover on the bottle.

Recorder sensor

 

 

(lower bottle of MBR-304G

 

Temp. sensor

and MBR-304GR)

 

Fixing screw

Cover

Fig. 1

Cap

 

 

 

Loosen

Fig. 2

13

Image 14
Contents MBR-304DR MBR-304GR MBR-304D MBR-304G Contents Precautions for Safe Operation Symbol showsPrecautions for Safe Operation Prevent electric shock or injury Damage Page Environmental Conditions Refrigerator Components Refrigerator Components Control panel and keypad Installation Installation siteInstallation Loosen Before Commencing OperationPreparation of monitor bottle Remote alarm terminal Operation after power failure START-UP of UnitAir-intake vent Operating InstructionsStoring of items Air-exhaust ventCycle defrost Temperature displayDefrost cycles Forced defrostCalibration of temperature sensor Alarm test procedure Alarms & Safety FunctionsAlarms and safety functions Replacement of fluorescent lamp Routine MaintenanceCleaning of cabinet Disposal of Unit Replacement of glow starterCleaning of condenser filter Alarm device is activated TroubleshootingIf nothing operates even when switched on When the unit does not get cold enoughSpecifications PerformanceSafety check sheet CEM