Sanyo MBR-304DR instruction manual Control panel and keypad

Page 11

REFRIGERATOR COMPONENTS

Control panel and keypad

 

7

 

 

 

8

 

 

 

9

 

 

10

 

 

 

 

 

 

 

 

 

 

 

11

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

1.Alarm buzzer stop key (BUZZER): To silence the audible alarm, press this key.

2.Temperature display key (UPPER/LOWER): By pressing this key, the displayed temperature is changed; upper monitor bottle temperature, lower monitor bottle temperature, and average temperature of upper and lower monitor bottles are displayed in sequence.

3.Alarm test key (ALARM TEST): To check the operation of alarm buzzer, lamp, and remote alarm terminal.

4.Calibration key (CAL): To calibrate the temperature. This key is used only when the temperature calibration is necessary.

5.Digit shift key ( ): Pressing this key in the CAL mode will cause the changeable digit to shift.

6.Numerical value shift key ( ): Pressing this key in the CAL mode will cause the numerical value to change.

7.Door check lamp (DOOR): This lamp is lit when the door is ajar or open.

8.Alarm lamp (ALARM): This lamp blinks during an alarm condition (temperature alarm or power failure alarm).

9, 10. Temperature display monitor lamp (UPPER, LOWER): Lamp 9 is lit when upper monitor bottle temperature is displayed. Lamp 10 is lit when lower monitor bottle temperature is displayed.

Both 9 and 10 are lit when the average temperature of lower and upper monitor bottle temperature is displayed.

11.Digital temperature indicator: This indicator shows the current chamber temperature.

10

Image 11
Contents MBR-304DR MBR-304GR MBR-304D MBR-304G Contents Symbol shows Precautions for Safe OperationPrecautions for Safe Operation Prevent electric shock or injury Damage Page Environmental Conditions Refrigerator Components Refrigerator Components Control panel and keypad Installation site InstallationInstallation Loosen Before Commencing OperationPreparation of monitor bottle Remote alarm terminal START-UP of Unit Operation after power failureAir-exhaust vent Operating InstructionsStoring of items Air-intake ventForced defrost Temperature displayDefrost cycles Cycle defrostCalibration of temperature sensor Alarm test procedure Alarms & Safety FunctionsAlarms and safety functions Replacement of fluorescent lamp Routine MaintenanceCleaning of cabinet Disposal of Unit Replacement of glow starterCleaning of condenser filter When the unit does not get cold enough TroubleshootingIf nothing operates even when switched on Alarm device is activatedPerformance SpecificationsSafety check sheet CEM