Characterizing a Filter

To display phase

Complete “To set up a frequency response measurement” or continue from one of the previous marker measurements.

1Display a second trace: Press [Display], [2 grids].

2Activate the second trace and define it as a frequency response measurement:

Press [B], [Measurement Data], [frequency response].

3Specify phase data for the second trace:

Press [Data Format], [phase wrap].

4Couple the markers on traces A and B: Press [Marker], [couple mkrs on].

5Move the markers with the knob to determine phase with respect to frequency response.

6Overlap the two traces:

Press [Shift], [A].

Press [Display] [single grid].

In this example, note that a trace which is displayed is not necessarily active (capable of being configured). You must specifically activate a displayed trace in order to change its configuration. For example, if you have chosen the relative marker in one trace then couple the markers, the marker on the second trace will be absolute, rather than relative, unless you activate the second trace and select the relative marker.

Coupling the markers on two traces lets you compare values at the same frequency

5-7

Image 39
Contents Agilent Technologies 89441A Getting Started Guide Analyzer at a Glance Front Panel IiiThis page left intentionally blank Saftey Summary Fuses Safety Symbols Notation Conventions This Book This page left intentionally blank Table of Contents Xii General TasksUsing Online Help Enter the online help system Press Help To learn about online helpTo display help for hardkeys and softkeys To display a related help topic To select a topic from the help index Page Making Simple Noise Measurements Start an averaged measurement To measure random noiseTurn on the band power markers To measure band powerTurn on the carrier-to-noise marker To measure signal to noise ratiosCarrier-to-noise ratio normalized to one Hertz Page Using Gating to Characterize a Burst Signal Display should now appear as shown below To Use Time GatingConfigure the display and the measurement Set up the time gating and examine the first burstExamine the second burst Page Measuring Relative Phase To measure the relative phase of an AM signal Measuring Relative Phase Zero the offset marker on the carrier To measure the relative phase of an PM signalCharacterizing a Filter To set up a frequency response measurement Have option AY7If section for network measurements Press Auto Scale To use the absolute marker To use the relative marker To use the search marker Specify phase data for the second trace To display phaseActivate the second trace and select a coherence measurement To display coherenceGeneral Tasks To set up peripherals General Tasks To save data with an internal or RAM disk Press Return To recall data with an internal or RAM diskTo format a disk To create a math function Define a constantDefine a math function Press Math, define F1 To use a math function Press measurement state or input/source state To display a summary of instrument parametersPage Preparing the Analyzer for Use Preparing the Analyzer for Use To chassis ground Analyzer cabinet can subject the operator to lethal voltages To do the incoming inspection Power if it is damagedPreparing the Analyzer for Use To connect the sections Preparing the Analyzer for Use To install the analyzer Select To change the if section’s line-voltage switchTo change the RF section’s line-voltage switch To change the if section’s fuse RF Section To change the RF section’s fuseInternet protocol address Return To connect the analyzer to a LANTo connect the analyzer to a parallel device To connect the analyzer to a serial deviceTo connect the analyzer to an external monitor To connect the analyzer to an Gpib deviceSet the if section’s power switch to on l To connect the optional keyboardConnect the other end of the keyboard cable to the keyboard To connect the optional minimum loss pad To store the analyzer To clean the screenElectricity which can damage electronic components To transport the analyzerIf the if section will not power up If the RF section will not power up If the analyzer’s stop frequency is 10 MHz Page Index Basic HT Gpib SNR HT Dqpsk HTQAM SNRSee also traces Gpib HT FSKGpib LAN Manuals, for this product HT See LAN RAM QAM HTSee knob RPGSource LED Tdma Sync not Found HTEXT Trigger LED HT PSD HTVSB Page Agilent 89400-Series Documentation Roadmap Need Assistance?