Emerson 3051S manual Filtering, Low Cutoff

Page 14

Rosemount 3051S

Reference Manual

00809-0200-4801, Rev BA April 2007

Filtering

The filtering feature changes the response time of the device to smooth

 

variations in output readings caused by rapid changes in input. Adjust the

 

filter time constant (in seconds) using the PV_FTIME parameter. Set the filter

 

time constant to zero to disable the filter feature.

Figure 3-3. Analog Input

PV_FTIME filtering Diagram.

Low Cutoff

OUT (mode in man)

OUT (mode in auto)

PV

63% of Change

FIELD_VAL

Time (seconds)

PV_FTIME

When the converted input value is below the limit specified by the LOW_CUT parameter, and the Low Cutoff I/O option (IO_OPTS) is enabled (True), a value of zero is used for the converted value (PV). This option is useful to eliminate false readings when the differential pressure measurement is close to zero, and it may also be useful with zero-based measurement devices such as flowmeters.

NOTE

Low Cutoff is the only I/O option supported by the AI block. Set the I/O option in Manual or Out of Service mode only.

3-14

Image 14
Contents Overview Safety Messages Reference ManualDevice Capabilities Link Active Scheduler Network Parameter ValueAuto General Block InformationModes MANBlock Instantiation Simulation FeaturesFeaturessel Resource BlockHwsel bit Swsel bit Write accessMaxnotify Read/WritePlantWeb Alarms Adviseenabled MaintalarmsMaintenabled MaintmaskConfigure the AI block Analog Input AI Function BlockChannel Number Channel Description ChannelLtype Pressure Channel Temperature Channel Flow Channel Configuration Examples Parameter Configured ValuesSituation #3 Diagram XDSCALE100 OUT= Filtering Low CutoffStatus Options Process AlarmsAlarm Priority Hipri Hihipri Lopri LolopriMultiple Analog Input MAI Function Block Advanced FeaturesAlarmtype OutdBlock Error Conditions Number Name and DescriptionCustom Meter Configuration LCD Transducer BlockPARAMINDEX#1 DisplayparamselBLKTYPE#1 CUSTOMTAG#1Custom Display bar graphCUSTOMUNITS#1 Advanced Diagnostics Transducer Block ADB ADB Parameters Plugged Impulse Line Detection PILConfiguration Plugged Impulse Line Detection PlineonPlugged Impulse Line Advanced Settings Rosemount 3051S Statistical Process Monitoring SPM SPM Configuration Mass Flow Installation and Setup Engineering Assistant SoftwareDevice View Open EA Wizard EA Wizard View 10. Download Mass Flow Configuration File
Related manuals
Manual 34 pages 49.8 Kb Manual 36 pages 50 Kb

3051S specifications

The Emerson 3051S is a state-of-the-art pressure transmitter that excels in various industrial applications. Its design is rooted in cutting-edge technology, ensuring accurate and reliable pressure measurements. With a focus on performance and durability, the 3051S stands as a trusted choice for engineers and process professionals worldwide.

One of the standout features of the Emerson 3051S is its advanced sensor technology. The device utilizes a high-performance sensor that offers exceptional accuracy, with a measured performance of up to ±0.075% of the span. This precision is essential for processes where minute changes in pressure can significantly impact operations. Coupled with the sensor's stability and repeatability, the 3051S provides consistent readings even in fluctuating conditions.

The device also incorporates Emerson's proprietary Smart Pressure technology, which enables predictive diagnostics. This feature allows users to anticipate potential issues before they escalate, reducing downtime and maintenance costs. The 3051S is equipped with advanced communication capabilities, supporting protocols such as HART, FOUNDATION Fieldbus, and Profibus PA. This versatility ensures seamless integration into existing plant automation systems, facilitating real-time data monitoring and management.

In terms of construction, the Emerson 3051S is built to withstand harsh industrial environments. It features a robust stainless-steel housing that protects the internal components from corrosive substances, high temperatures, and extreme pressures. The device is also available with various membrane materials, making it suitable for a wide range of media, including corrosive and viscous fluids.

Moreover, the 3051S is designed with user-friendliness in mind. It includes an intuitive interface and is fully configurable, allowing for easy calibration and setup. The ability to use Emerson's AMS Device Manager software further enhances the user experience, enabling detailed analysis, configuration, and diagnostics from a centralized platform.

Overall, the Emerson 3051S combines superior technology with practical design, making it an ideal solution for industries such as oil and gas, chemical processing, and water treatment. Its precise measurements, advanced diagnostics, and durable construction position it as a benchmark in the field of pressure transmitters, delivering reliability and efficiency for critical applications.