Emerson 3051S manual Maintalarms, Maintenabled, Maintmask, Maintpri, Maintactive, Maintalm

Page 7

Reference Manual

00809-0200-4801, Rev BA April 2007

Rosemount 3051S

MAINT_ALARMS

A maintenance alarm indicates the device or some part of the device needs maintenance soon. If the condition is ignored, the device will eventually fail. There are five parameters associated with MAINT_ALARMS, they are described below.

MAINT_ENABLED

The MAINT_ENABLED parameter contains a list of conditions indicating the device or some part of the device needs maintenance soon. If the condition is ignored, the device will eventually fail.

Below is a list of the conditions with the highest priority first.

1.Sensor Module Memory Warning

2.Primary Value Degraded

3.Secondary Value Degraded

4.Plugged Impulse Line Detected

MAINT_MASK

The MAINT_MASK parameter will mask any of the failed conditions listed in MAINT_ENABLED. A bit on means that the condition is masked out from alarming and will not be reported.

MAINT_PRI

MAINT_PRI designates the alarming priority of the MAINT_ALM, “Process Alarms” on page 3-15. The default is 0 and the recommended values is 3 to 7.

MAINT_ACTIVE

The MAINT_ACTIVE parameter displays which of the alarms is active. Only the condition with the highest priority will be displayed. This priority is not the same as the MAINT_PRI parameter described above. This priority is hard coded within the device and is not user configurable.

MAINT_ALM

An alarm indicating the device needs maintenance soon. If the condition is ignored, the device will eventually fail.

Advisory Alarms

An advisory alarm indicates informative conditions that do not have a direct impact on the device's primary functions There are five parameters associated with ADVISE_ALARMS, they are described below.

ADVISE_ENABLED

The ADVISE_ENABLED parameter contains a list of informative conditions that do not have a direct impact on the device's primary functions. Below is a list of the advisories with the highest priority first.

3-7

Image 7
Contents Reference Manual Overview Safety MessagesNetwork Parameter Value Device Capabilities Link Active SchedulerMAN General Block InformationModes AutoResource Block FeaturesFeaturessel Block Instantiation SimulationRead/Write Write accessMaxnotify Hwsel bit Swsel bitPlantWeb Alarms Maintmask MaintalarmsMaintenabled AdviseenabledAnalog Input AI Function Block Configure the AI blockLtype ChannelChannel Number Channel Description Pressure Channel Temperature Channel Flow Channel Parameter Configured Values Configuration ExamplesSituation #3 Diagram XDSCALE100 OUT= Low Cutoff FilteringHipri Hihipri Lopri Lolopri Process AlarmsAlarm Priority Status OptionsOutd Advanced FeaturesAlarmtype Multiple Analog Input MAI Function BlockNumber Name and Description Block Error ConditionsLCD Transducer Block Custom Meter ConfigurationCUSTOMTAG#1 DisplayparamselBLKTYPE#1 PARAMINDEX#1CUSTOMUNITS#1 Display bar graphCustom ADB Parameters Plugged Impulse Line Detection PIL Advanced Diagnostics Transducer Block ADBPlineon Configuration Plugged Impulse Line DetectionPlugged Impulse Line Advanced Settings Rosemount 3051S Statistical Process Monitoring SPM SPM Configuration Mass Flow Engineering Assistant Software Installation and SetupDevice View Open EA Wizard EA Wizard View 10. Download Mass Flow Configuration File
Related manuals
Manual 34 pages 49.8 Kb Manual 36 pages 50 Kb

3051S specifications

The Emerson 3051S is a state-of-the-art pressure transmitter that excels in various industrial applications. Its design is rooted in cutting-edge technology, ensuring accurate and reliable pressure measurements. With a focus on performance and durability, the 3051S stands as a trusted choice for engineers and process professionals worldwide.

One of the standout features of the Emerson 3051S is its advanced sensor technology. The device utilizes a high-performance sensor that offers exceptional accuracy, with a measured performance of up to ±0.075% of the span. This precision is essential for processes where minute changes in pressure can significantly impact operations. Coupled with the sensor's stability and repeatability, the 3051S provides consistent readings even in fluctuating conditions.

The device also incorporates Emerson's proprietary Smart Pressure technology, which enables predictive diagnostics. This feature allows users to anticipate potential issues before they escalate, reducing downtime and maintenance costs. The 3051S is equipped with advanced communication capabilities, supporting protocols such as HART, FOUNDATION Fieldbus, and Profibus PA. This versatility ensures seamless integration into existing plant automation systems, facilitating real-time data monitoring and management.

In terms of construction, the Emerson 3051S is built to withstand harsh industrial environments. It features a robust stainless-steel housing that protects the internal components from corrosive substances, high temperatures, and extreme pressures. The device is also available with various membrane materials, making it suitable for a wide range of media, including corrosive and viscous fluids.

Moreover, the 3051S is designed with user-friendliness in mind. It includes an intuitive interface and is fully configurable, allowing for easy calibration and setup. The ability to use Emerson's AMS Device Manager software further enhances the user experience, enabling detailed analysis, configuration, and diagnostics from a centralized platform.

Overall, the Emerson 3051S combines superior technology with practical design, making it an ideal solution for industries such as oil and gas, chemical processing, and water treatment. Its precise measurements, advanced diagnostics, and durable construction position it as a benchmark in the field of pressure transmitters, delivering reliability and efficiency for critical applications.