Emerson 3051S manual PlantWeb Alarms

Page 6

Rosemount 3051S

Reference Manual

00809-0200-4801, Rev BA April 2007

PlantWebAlarms

The alarms and recommended actions should be used in conjunction with

 

Section 2: Troubleshooting.

 

The Resource Block will act as a coordinator for PlantWeb alarms. There will

 

be three alarm parameters (FAILED_ALARM, MAINT_ALARM, and

 

ADVISE_ALARM) which will contain information regarding some of the device

 

errors which are detected by the transmitter software. There will be a

 

RECOMMENDED_ACTION parameter which will be used to display the

 

recommended action text for the highest priority alarm. FAILED_ALARM will

 

have the highest priority followed by MAINT_ALARM and ADVISE_ALARM

 

will be the lowest priority.

 

 

 

NOTE

 

PlantWeb alerts for ADB and mass flow cannot be simulated.

 

 

 

FAILED_ALARMS

 

A failure alarm indicates a failure within a device that will make the device or

 

some part of the device non-operational. This implies that the device is in

 

need of repair and must be fixed immediately. There are five parameters

 

associated with FAILED_ALARMS specifically, they are described below.

 

FAILED_ENABLED

 

This parameter contains a list of failures in the device which makes the

 

device non-operational that will cause an alarm to be sent. Below is a list

 

of the failures with the highest priority first.

 

1.

Memory Failure

 

2.

NV Memory Failure

 

3.

Primary Value Failure

 

4.

Secondary Value Failure

 

5.

Sensor Module Memory Failure

 

6.

Sensor Module Failure

FAILED_MASK

This parameter will mask any of the failed conditions listed in

FAILED_ENABLED. A bit on means that the condition is masked out from alarming and will not be reported.

FAILED_PRI

Designates the alarming priority of the FAILED_ALM, see “Alarm Priority” on page 3-15. The default is 0 and the recommended value are between 8 and 15.

FAILED_ACTIVE

This parameter displays which of the alarms is active. Only the alarm with the highest priority will be displayed. This priority is not the same as the FAILED_PRI parameter described above. This priority is hard coded within the device and is not user configurable.

FAILED_ALM

Alarm indicating a failure within a device which makes the device non-operational.

3-6

Image 6
Contents Overview Safety Messages Reference ManualDevice Capabilities Link Active Scheduler Network Parameter Value Auto General Block Information Modes MANBlock Instantiation Simulation FeaturesFeaturessel Resource BlockHwsel bit Swsel bit Write accessMaxnotify Read/WritePlantWeb Alarms Adviseenabled MaintalarmsMaintenabled MaintmaskConfigure the AI block Analog Input AI Function BlockChannel LtypeChannel Number Channel Description Pressure Channel Temperature Channel Flow Channel Configuration Examples Parameter Configured ValuesSituation #3 Diagram XDSCALE100 OUT= Filtering Low CutoffStatus Options Process AlarmsAlarm Priority Hipri Hihipri Lopri LolopriMultiple Analog Input MAI Function Block Advanced FeaturesAlarmtype OutdBlock Error Conditions Number Name and DescriptionCustom Meter Configuration LCD Transducer BlockPARAMINDEX#1 DisplayparamselBLKTYPE#1 CUSTOMTAG#1Display bar graph CUSTOMUNITS#1Custom Advanced Diagnostics Transducer Block ADB ADB Parameters Plugged Impulse Line Detection PILConfiguration Plugged Impulse Line Detection PlineonPlugged Impulse Line Advanced Settings Rosemount 3051S Statistical Process Monitoring SPM SPM Configuration Mass Flow Installation and Setup Engineering Assistant SoftwareDevice View Open EA Wizard EA Wizard View 10. Download Mass Flow Configuration File
Related manuals
Manual 34 pages 49.8 Kb Manual 36 pages 50 Kb

3051S specifications

The Emerson 3051S is a state-of-the-art pressure transmitter that excels in various industrial applications. Its design is rooted in cutting-edge technology, ensuring accurate and reliable pressure measurements. With a focus on performance and durability, the 3051S stands as a trusted choice for engineers and process professionals worldwide.

One of the standout features of the Emerson 3051S is its advanced sensor technology. The device utilizes a high-performance sensor that offers exceptional accuracy, with a measured performance of up to ±0.075% of the span. This precision is essential for processes where minute changes in pressure can significantly impact operations. Coupled with the sensor's stability and repeatability, the 3051S provides consistent readings even in fluctuating conditions.

The device also incorporates Emerson's proprietary Smart Pressure technology, which enables predictive diagnostics. This feature allows users to anticipate potential issues before they escalate, reducing downtime and maintenance costs. The 3051S is equipped with advanced communication capabilities, supporting protocols such as HART, FOUNDATION Fieldbus, and Profibus PA. This versatility ensures seamless integration into existing plant automation systems, facilitating real-time data monitoring and management.

In terms of construction, the Emerson 3051S is built to withstand harsh industrial environments. It features a robust stainless-steel housing that protects the internal components from corrosive substances, high temperatures, and extreme pressures. The device is also available with various membrane materials, making it suitable for a wide range of media, including corrosive and viscous fluids.

Moreover, the 3051S is designed with user-friendliness in mind. It includes an intuitive interface and is fully configurable, allowing for easy calibration and setup. The ability to use Emerson's AMS Device Manager software further enhances the user experience, enabling detailed analysis, configuration, and diagnostics from a centralized platform.

Overall, the Emerson 3051S combines superior technology with practical design, making it an ideal solution for industries such as oil and gas, chemical processing, and water treatment. Its precise measurements, advanced diagnostics, and durable construction position it as a benchmark in the field of pressure transmitters, delivering reliability and efficiency for critical applications.