Emerson Comprehensive Guide to the 3051S Manual and Configuration

Page 32

Rosemount 3051S

Reference Manual

00809-0200-4801, Rev BA April 2007

Figure 3-10. Download Mass

Flow Configuration File

Selected

 

 

Device

Send Mass Flow

 

(Highlighted)

 

Configuration File to

Status Confirms Mass

 

 

Selected Device

 

Flow Configuration File

 

 

 

 

Installations

Table 3-7. Mass Flow Block Device - Specific Parameter Values

 

 

Data Type/

 

 

 

 

 

 

 

Index

Parameter

Structure

Store

Size

Valid Range

Initial Value

Units

Mode

Description

 

 

 

 

 

 

 

 

 

 

14

DIFFERENTIAL

DS-65

D

5

 

 

inH20

MAN

The differential pressure status and

 

_PRESSURE

 

 

 

 

 

@68°F

 

value

 

 

 

 

 

 

 

 

 

 

15

DIFFERENTIAL

Unsigned16

S

2

10: DP Sensor

10

E

O/S

The channel through which the

 

_PRESSURE_

 

 

 

(scaled)

 

 

 

differential pressure is communicated

 

SOURCE

 

 

 

 

 

 

 

(and converted to the proper units)

16

PRESSURE

DS-65

D

5

 

 

Psia

MAN,

The absolute pressure status and value

 

 

 

 

 

 

 

 

O/S(1)

 

17

PRESSURE_

Unsigned16

S

2

5: Absolute

8

E

O/S

The channel through which the absolute

 

SOURCE

 

 

 

Pressure

 

 

 

pressure is communicated. When

 

 

 

 

 

(A0.OUT)

 

 

 

AO.OUT value is used, FAULT_STATE

 

 

 

 

 

8: Absolute

 

 

 

is used if status goes bad. When

 

 

 

 

 

Pressure

 

 

 

CAS_IN shadow is used, bad

 

 

 

 

 

(CAS_IN

 

 

 

status/value will be propagated to the

 

 

 

 

 

shadow)

 

 

 

mass flow block. If a value of “Constant”

 

 

 

 

 

255: Constant

 

 

 

is specified, the PRESSURE parameter

 

 

 

 

 

 

 

 

 

becomes writable in O/S and the

 

 

 

 

 

 

 

 

 

constant value is used in the calculation.

18

TEMPERATUR

DS-65

D

5

 

 

°F

MAN,

The process temperature status and

 

E

 

 

 

 

 

 

O/S(2)

value

19

TEMPERATUR

Unsigned16

S

2

6: Process

9

E

O/S

The channel through which the process

 

E_SOURCE

 

 

 

Temperature

 

 

 

temperature is communicated. When

 

 

 

 

 

(CAS_IN

 

 

 

AO.OUT value is used, FAULT_STATE

 

 

 

 

 

shadow)

 

 

 

is used if status goes bad. When

 

 

 

 

 

9: Process

 

 

 

CAS_IN shadow is used, bad

 

 

 

 

 

Temperature

 

 

 

status/value will be propagated to the

 

 

 

 

 

(CAS_IN

 

 

 

mass flow block. If a value of “Constant”

 

 

 

 

 

shadow)

 

 

 

is specified, the TEMPERATURE

 

 

 

 

 

255: Constant

 

 

 

parameter becomes writable in O/S and

 

 

 

 

 

 

 

 

 

the constant value is used in the

 

 

 

 

 

 

 

 

 

calculation.

 

 

 

 

 

 

 

 

 

 

(1)PRESSURE can be written in O/S mode if the PRESSURE_SOURCE is set to “Constant”

(2)TEMPERATURE can be written in O/S mode if the TEMPERATURE_SOURCE is set to “Constant”

3-32

Image 32
Contents Overview Safety Messages Reference ManualDevice Capabilities Link Active Scheduler Network Parameter ValueGeneral Block Information ModesAuto MANFeatures FeaturesselBlock Instantiation Simulation Resource BlockWrite access MaxnotifyHwsel bit Swsel bit Read/WritePlantWeb Alarms Maintalarms MaintenabledAdviseenabled MaintmaskConfigure the AI block Analog Input AI Function BlockChannel Number Channel Description ChannelLtype Pressure Channel Temperature Channel Flow Channel Configuration Examples Parameter Configured ValuesSituation #3 Diagram XDSCALE100 OUT= Filtering Low CutoffProcess Alarms Alarm PriorityStatus Options Hipri Hihipri Lopri LolopriAdvanced Features AlarmtypeMultiple Analog Input MAI Function Block OutdBlock Error Conditions Number Name and DescriptionCustom Meter Configuration LCD Transducer BlockDisplayparamsel BLKTYPE#1PARAMINDEX#1 CUSTOMTAG#1Custom Display bar graphCUSTOMUNITS#1 Advanced Diagnostics Transducer Block ADB ADB Parameters Plugged Impulse Line Detection PILConfiguration Plugged Impulse Line Detection PlineonPlugged Impulse Line Advanced Settings Rosemount 3051S Statistical Process Monitoring SPM SPM Configuration Mass Flow Installation and Setup Engineering Assistant Software Device View Open EA Wizard EA Wizard View 10. Download Mass Flow Configuration File
Related manuals
Manual 34 pages 49.8 Kb Manual 36 pages 50 Kb

3051S specifications

The Emerson 3051S is a state-of-the-art pressure transmitter that excels in various industrial applications. Its design is rooted in cutting-edge technology, ensuring accurate and reliable pressure measurements. With a focus on performance and durability, the 3051S stands as a trusted choice for engineers and process professionals worldwide.

One of the standout features of the Emerson 3051S is its advanced sensor technology. The device utilizes a high-performance sensor that offers exceptional accuracy, with a measured performance of up to ±0.075% of the span. This precision is essential for processes where minute changes in pressure can significantly impact operations. Coupled with the sensor's stability and repeatability, the 3051S provides consistent readings even in fluctuating conditions.

The device also incorporates Emerson's proprietary Smart Pressure technology, which enables predictive diagnostics. This feature allows users to anticipate potential issues before they escalate, reducing downtime and maintenance costs. The 3051S is equipped with advanced communication capabilities, supporting protocols such as HART, FOUNDATION Fieldbus, and Profibus PA. This versatility ensures seamless integration into existing plant automation systems, facilitating real-time data monitoring and management.

In terms of construction, the Emerson 3051S is built to withstand harsh industrial environments. It features a robust stainless-steel housing that protects the internal components from corrosive substances, high temperatures, and extreme pressures. The device is also available with various membrane materials, making it suitable for a wide range of media, including corrosive and viscous fluids.

Moreover, the 3051S is designed with user-friendliness in mind. It includes an intuitive interface and is fully configurable, allowing for easy calibration and setup. The ability to use Emerson's AMS Device Manager software further enhances the user experience, enabling detailed analysis, configuration, and diagnostics from a centralized platform.

Overall, the Emerson 3051S combines superior technology with practical design, making it an ideal solution for industries such as oil and gas, chemical processing, and water treatment. Its precise measurements, advanced diagnostics, and durable construction position it as a benchmark in the field of pressure transmitters, delivering reliability and efficiency for critical applications.