Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.

www.rangerfluidwarming.com reorder #202456B

English

3MTM RangerTM Irrigation Fluid Warming System

 

 

WARNING:

1.To reduce the risks associated with hazardous voltage and fire and thermal energy hazards:

Do not substitute other devices (i.e. Model 245) for the Ranger irrigation fluid warming unit or Ranger irrigation disposable sets.

Do not continue use of the unit if the over-temperature alert continues to sound and the temperature does not return to the set point temperature. Immediately stop fluid flow and discard the disposable set. Have the warming unit tested by a biomedical technician or call 3M Patient Warming technical service.

2.To reduce the risks associated with hazardous voltage and fire:

Do not modify or service this device, and do not open the warming unit case- there are no user- serviceable parts in the Ranger warming unit.

Connect power cord to receptacles marked “Hospital Only,” “Hospital Grade,” or a reliably grounded outlet.

Use only the power cord specified for this product and certified for the country of use.

Do not allow the power cord to get wet.

Do not use the Ranger warming system when it appears the warming unit, power cord, or disposable is damaged. Use only 3M specified replacement parts.

Keep power cord visible and accessible at all times. The plug on the power cord serves as the disconnect device. The wall socket outlet shall be as close as practical and shall be easily accessible.

3.To reduce the risks associated with air embolism:Never use fluids if air bubbles are present in the fluid line.

CAUTION:

1.Not for I.V. use.2.To reduce the risks associated with cross-contamination:

The cleaning tool provides only superficial cleaning, it does not disinfect or sterilize the interior of the unit.

3.To reduce the risks associated with impact and facility medical device damage:

Clamp the Ranger irrigation fluid warming unit to an I.V. pole with a minimum 14” (35.6 cm) radius wheelbase and at a height no higher than 44” (112 cm).

4.To reduce the risks associated with environmental contamination:

Follow applicable regulations when disposing of this device or any of its electronic components.

NOTICE:

1.Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.

2.To avoid device damage:

Do not clean the warming unit with solvents. Damage to the case, label, and internal components may result.

Do not immerse the warming unit in cleaning or sterilizing solutions. The unit is not liquid proof.

Do not insert metallic instruments in the warming unit.Do not use abrasive materials or solutions to clean the heater plates.

Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit.

3.The Ranger irrigation fluid warming unit meets medical electronic interference requirements. If radio frequency interference with other equipment should occur, connect the unit to a different power source.

6