Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.

www.rangerfluidwarming.com reorder #202456B

English

3MTM RangerTM Irrigation Fluid Warming System

 

 

Model 247 Product Safety Features

The following chart describes the safety alarm features of the Ranger irrigation fluid warming unit.

Alert Type

 

Set-Point

 

Indication

Primary over-temperature

 

48°C

 

LED alternately displays

 

 

alert system

 

 

 

“HI” and temperature,

 

 

 

 

alert sounds, alert

 

 

 

 

light flashes.

Independent back-up

 

50°C

 

LED displays the

over-temperature alert

 

 

 

temperature or has gone

system (cut-off).

 

 

 

dark, alert sounds.

Under-temperature alert

 

33°C

 

LED displays “LO” and

 

 

 

 

temperature, alert sounds,

 

 

 

 

alert light flashes.

 

 

 

 

 

Description

Primary Alert: Alert sounds, controller cuts power to the heaters until temperature falls below 48°C. When heater temperature drops below 48°C, audible and visual alerts stop.

Primary failure. Temperature will rise to 50°C within a few seconds. The safety backup system is activated at 50°C and power is immediately shut off to the heating plates.

Heater temperature has fallen to 33°C. Alerts will stop when temperature rises above 33°C. Continue use of unit.

NOTE:

The difference in over-temperature alert system cut-off points does not indicate any difference in safety and is simply a result of tighter controller tolerances.

The Ranger irrigation fluid warming system is designed with very tight controls to regulate its temperature to the 41°C set-point. The microprocessor controller measures the temperature at the fluid pathway inside the warming plate four times every second, making it very responsive to changes in flow rate.

8