Check the 3MTM RangerTM system website to ensure you have the most recent version of this document.

www.rangerfluidwarming.com reorder #202456B

English

3MTM RangerTM Irrigation Fluid Warming System

 

 

Specifications

Physical Characteristics

warming unit

4.5in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)

certifications

IEC/EN 60601-1; CAN/CSA-C22.2, No.601.1

classification

MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No.4HZ8

Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B, Ordinary equipment, Continuous operation. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only, in accordance with IEC/EN 60601-1 and in accordance with Canadian/ CSA C22.2, No. 601.1. Classified under the Medical Device Directive as a Class IIb device.

Electrical characteristics

Temperature characteristics

input voltage

set point temperature

100-120 or 220-240 VAC

41°C

operating frequency

over-temperature alert

100-120 VAC, 50/60 Hz

48°C

220-240 VAC, 50/60 Hz

 

maximum heating power

under-temperature alert

900 W

33°C

fuse

over-temperature cutoff

2 x T10A-H (250V) for 100-120 VAC

50°C

2 x T6.3A-H (250V) for 220-240 VAC

 

fuse type

 

Time delay, high breaking

 

Leakage Current

Meets leakage current requirements in accordance with UL/IEC 60601-1.

Environmental Conditions

operating temperature range

15° to 40°C (59° to 104°F)

storage temperature range

-20° to 45°C (-4° to 113°F)

operating humidity

10 to 85% RH, noncondensing

atmospheric pressure range

50 kPa to 106 kPa

14