Check the 3MTM RangerTM system website to ensure you have the most recent version of this document. | www.rangerfluidwarming.com reorder #202456B |
3MTM RangerTM Irrigation Fluid Warming System | English |
|
|
Condition | Cause |
Alert sounds but warming | Independent backup safety |
unit has been turned off. | system has been activated. |
Condition | Cause |
Cannot remove cassette from unit. | • Cassette is too full, fluids |
| are still being infused, or |
| clamp is open proximal to |
| the warming cassette. |
| • Warming unit is below |
| patient level, creating |
| excessive back pressure. |
Solution
Unplug unit. Call 3M Patient Warming technical support.
Solution
•Make sure fluid is drained from warming cassette before sliding out the cassette, that fluids are no longer infusing, and that clamp is closed proximal to the warming cassette.
•Raise unit above patient level.
Condition
Alerts indicator light illuminates and alert sounds,
Cause
Under temperature condition caused by very high flow using very cold fluid, or defective heater/relay.
Solution
Alerts should stop when temperature rises above 33°C. If alerts continue, turn off unit, unplug unit, and discontinue use. Call 3M Patient Warming technical support.
Condition | Cause | Solution |
Open wire on temperature sensor. | |
reads “Er 4” or “Open.” |
|
Do not use unit. Call 3M Patient Warming technical support.
Condition | Cause | Solution |
Electrical interference. | Remove the unit. Refer to biomedical | |
reads “Er 5” or “Open.” |
| technician or call 3M Patient |
|
| Warming technical support. |
202456B English | 11 |
|