English | 3MTM RangerTM Irrigation Fluid Warming System |
|
|
Troubleshooting
Condition | Cause |
Nothing illuminates on the | • Unit is not turned on, plugged |
warming unit panel. | in, or power cord is not plugged |
| into an appropriate outlet. |
| • Unit failure. |
•Turn unit on. Make sure the power cord is plugged into the power entry module of the warming unit. Make sure the warming unit is plugged into a properly grounded outlet.
•Call 3M Patient Warming technical support.
Condition
Alert indicator illuminates and alert sounds,
Cause
Temporary
•An extreme change in flow rates occurred (e.g., from
500 ml/min to stop flow).
•Unit was turned on and reached set point temperature before warming cassette was inserted.
•Fluids were prewarmed to above 42°C before being run through the warming unit.
Solution
•Open flow to reduce temperature. Alerts will stop when the temperature drops below 48°C. The unit is ready to use.
•Alerts will stop when the temperature drops below 48°C. The unit is ready to use.
•Turn off unit and unplug it. Discontinue infusion of fluids. Do not warm fluids before infusing them through the Ranger irrigation fluid warming unit.
Condition | Cause | Solution |
Alert sounds, | Primary controller failure. | Power to heating plates will shut off |
and alert indicator light go dark. | Unit will no longer operate. | if temperature rises to 50°C. Turn |
|
| unit off and unplug it. Discontinue |
|
| use of unit. Discard disposable set. |
|
| Alert will continue to sound if you |
|
| do not unplug unit. Call 3M Patient |
|
| Warming technical support. |
Condition | Cause |
Unit alerts soon after plugging it | Test screw on bottom of |
in (unit does not have to be turned | unit is loose or missing. |
on for this condition to occur). |
|
Heater temperature rises to 50°C and |
|
unit shuts down soon after plugging |
|
it in (unit does not have to be turned |
|
on for this condition to occur). |
|
Solution
Make sure test screw is completely tightened. If it is missing, turn unit off and unplug it. Call 3M Patient Warming technical support.