Error

Meaning

Status

 

Operation and Treatment

 

code

 

 

 

 

 

E30

Measurement is

Air is exhausted from the

 

Repair is necessary because of slow

 

greater than 120

cuff, and an error code is

 

inflation or slow constant exhaust.

 

seconds.

displayed.

 

 

 

 

 

 

 

E31

The constant

Air is exhausted from the

 

Repair is necessary because of slow

 

exhaust is greater

cuff, and an error code is

 

constant exhaust.

 

than 60 seconds.

displayed.

 

 

 

 

 

 

 

E32

Clock error.

An error code is displayed.

 

If you not clear this error, the product

 

 

 

 

needs repair.

 

 

 

 

 

E50

Pressure offset

An error code is displayed

 

Release the air from the cuff com-

 

error to measure

at restarting the product.

 

pletely, reset the product. If you not

 

pulsation.

 

 

clear this error, repair is necessary.

 

 

 

 

 

E52

Memory error.

An error code is displayed

 

The product needs repair.

 

 

at restarting the product.

 

 

 

 

 

 

 

E53

Battery contact is

The measurement is

 

Replace batteries correctly. If you can

 

defective.

stopped, air is released

 

not clear this error, the product needs

 

 

from the cuff and an error

 

repair.

 

 

code is displayed.

 

 

 

 

 

 

 

E55

Exhaust error.

An error code is displayed

 

Relax and do not move during the

E56

 

at measurement.

 

measurement. If this error occurs

E57

 

 

 

many times, repair is necessary.

 

 

 

 

 

E60

Interval setting

Start time is not proper,

 

Enter parameters for the interval

 

error.

interval of last block is not

 

correctly.

 

 

set in the unit of 120 min.

 

 

 

 

 

 

 

E70

RS-232C error.

The error code is displayed

 

Re-connect the communication cable.

E71

 

during communications.

 

If you can not clear this error, the

E72

 

 

 

product needs repair.

E73

 

 

 

 

 

 

 

 

 

E74

Low battery for

 

 

Replace batteries with new ones and

 

communication.

 

 

restart communication.

 

 

 

 

 

E75

Protcol error due

 

 

Re-connect the communication cable.

 

to external

 

 

If you can not clear this error, the

 

equipment.

 

 

product needs repair.

 

 

 

 

 

E90

Pressure zero

This error code is

 

Release the air from the cuff

 

error for safety

displayed before the

 

completely.

 

circuit.

measurement.

 

 

 

 

 

 

 

E91

Safety circuit

Patient moved during the

 

Relax and try to quiet during the

 

detects over load

measurement.

 

measurement. If it occurs in quiet, the

 

pressure.

 

 

recorder needs repair.

 

 

 

 

 

Other

 

Monitor code is displayed.

 

Reset. Turn on power switch again.

 

 

 

 

 

Maintenance

Page 33

Error Code

Page 35
Image 35
A&D Ambulatory Blood Pressure Monitor instruction manual Maintenance Error Code

Ambulatory Blood Pressure Monitor specifications

The A&D Ambulatory Blood Pressure Monitor is a cutting-edge device designed for the accurate measurement of blood pressure over an extended period. This innovative tool is particularly beneficial for diagnosing hypertension and evaluating blood pressure variability, making it essential for both healthcare professionals and patients.

One of the main features of the A&D Ambulatory Blood Pressure Monitor is its ability to perform measurements during the patient’s normal daily activities. This is crucial for capturing accurate readings, as blood pressure can fluctuate significantly throughout the day due to various factors such as stress, physical activity, and diet. By documenting blood pressure readings at set intervals over 24 hours, this monitor provides a comprehensive overview of the patient's cardiovascular health.

The device utilizes advanced oscillometric technology, which enhances accuracy by analyzing the oscillations in the arterial wall during cuff inflation. This method is recognized for its reliability compared to traditional auscultatory techniques. The monitor is equipped with a large memory capacity, allowing it to store numerous readings which can be analyzed later. This feature is invaluable for doctors who need to review the patient’s blood pressure patterns over time.

User-friendliness is another hallmark of the A&D Ambulatory Blood Pressure Monitor. The device typically features an intuitive display and controls that make it easy to navigate through settings and review stored data. Many models offer wireless connectivity, enabling seamless data transfer to smartphones or computers for further analysis and reporting. This capability aids healthcare providers in timely decision-making regarding treatment options.

Additionally, the A&D monitor is designed with comfort in mind. It comes with a soft, adjustable cuff that can accommodate a range of arm sizes, ensuring an accurate fit and comfortable experience for the patient. The device is ergonomically designed and lightweight, making it easy to wear throughout the day without interfering with daily activities.

The A&D Ambulatory Blood Pressure Monitor not only promotes better patient compliance but also enhances clinical outcomes by enabling continuous and accurate monitoring of blood pressure in real-world settings. Overall, this device embodies the convergence of technology and healthcare, providing invaluable data that fosters informed decisions and improves patient care.