EMC TABLE

Guidance and manufacturer’s declaration – electromagnetic immunity

The A&D unit is intended for use in the electromagnetic environment specified below. The customer or the user of the A&D unit should assure that it is used in such an environment.

Immunity

IEC 60601

Compliance

Electromagnetic environment –

test

test level

level

guidance

 

 

 

 

 

 

 

Portable and mobile RF

 

 

 

communications equipment should

 

 

 

be used no closer to any part of

 

 

 

the A&D unit, including cables,

 

 

 

than the recommended separation

 

 

 

distance calculated from the

 

 

 

equation applicable to the

 

 

 

frequency of the transmitter.

 

 

 

Recommended separation distance:

Conducted RF

3 V rms

3 V rms

d = 1.2

P

 

IEC 61000-4-6 150 kHz to 80 MHz

 

 

 

 

 

Radiated RF

3 V/m

3 V/m

d = 1.2

P

80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2,5 GHz

d = 2.3

P

800 MHz to 2,5 GHz

 

 

 

 

 

 

where P is the maximum output

 

 

 

power rating of the transmitter in

 

 

 

watts (W) according to the

 

 

 

transmitter manufacturer and d is

 

 

 

the recommended separation

 

 

 

distance in metres (m).

 

 

 

Field strengths from fixed RF

 

 

 

transmitters, as determined by an

 

 

 

electromagnetic site survey, a

 

 

 

should be less than the compliance

 

 

 

level in each frequency range.b

 

 

 

Interference may occur in

 

 

 

the vicinity

of equipment

 

 

 

marked with the following

 

 

 

symbol:

 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the A&D unit is used exceeds the applicable RF compliance level above, the A&D unit should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re- orienting or relocating the A&D unit.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991

E-28

Page 29
Image 29
A&D UA-1020CN instruction manual Recommended separation distance