TROUBLESHOOTING

Problem

Probable Cause

Corrective Action

 

 

 

Nothing appears

Batteries are drained.

Replace all batteries with

in the display, after

 

new ones. Alkaline batteries

I press the START

 

recommended.

button.

 

 

Battery terminals are not in

Reinstall the batteries with

 

 

the correct position.

negative and positive termi-

 

 

nals matching those indicated

 

 

in the battery compartment.

The cuff does not

Battery voltage is too low.

Replace all batteries with

inflate.

Low battery symbol blinks.

new ones. Alkaline batteries

 

[If the batteries are drained

recommended.

 

completely, the mark does

 

 

not appear.]

 

 

 

 

The unit does not

The cuff is not fastened

Fasten the cuff correctly.

measure. Readings

properly.

 

are too high or too

 

 

You moved your arm or

Make sure you remain very

low.

body during the

still and quiet during the

 

 

measurement.

measurement.

 

 

 

 

The cuff position is not

Sit comfortably and still.

 

correct.

Make sure the cuff is at the

 

 

same level as your heart.

 

You are using the wrong

See Pg. E-9 "Selecting the

 

size cuff.

Correct Cuff."

 

 

 

 

Taking too many readings

Relax several minutes before

 

on the same arm in a short

 

each measurement.

 

period of time.

 

 

 

 

 

The value is

The healthcare professional

See Pg. E-23 "Answers to

different from that

may be using a different

Why Your Readings Are

measured at a clinic

sized cuff.

Different Between Home and

or doctor’s office.

 

the Doctor’s Office."

 

 

 

 

Your measurements may

See Pg. E-22 "Why Measure

 

be elevated by white coat

Blood Pressure at Home."

 

hypertension

 

 

 

 

NOTE: If the actions described above do not solve the problem, call 1-888-726-9966. Do not attempt to repair the device yourself.

E-25

A&D Medical (Toll-Free): 1-888-726-9966

Page 26
Image 26
A&D UA-1020CN instruction manual Troubleshooting, Problem Probable Cause Corrective Action