EMC TABLE
Guidance and manufacturer’s declaration – electromagnetic immunity
The A&D unit is intended for use in the electromagnetic environment specified below. The customer or the user of the A&D unit should assure that it is used in such an environment.
Immunity | IEC 60601 | Compliance | Electromagnetic |
| ||||
test | test level | level | environment – guidance | |||||
|
|
|
| |||||
Electrostatic | ± 6 kV | ± 6 kV | Floors should be wood, | |||||
discharge (ESD) | contact | contact | concrete or ceramic tile. | |||||
IEC | ± 8 kV | ± 8 kV | If floors are covered with | |||||
| synthetic material, the | |||||||
| air | air | ||||||
| relative | humidity | should | |||||
|
|
| ||||||
|
|
| be at least 30%. |
| ||||
|
|
|
| |||||
Electrical fast | ± 2 kV for power | ± 2 kV for power | Mains power quality | |||||
transient/burst | supply lines | supply lines | should be that of a typical | |||||
IEC | ± 1 kV | ± 1 kV | commercial | or | hospital | |||
| environment. |
|
| |||||
| for input/output | for input/output |
|
| ||||
|
|
|
|
|
| |||
| lines | lines |
|
|
|
|
| |
Surge | ± 1 kV | ± 1 kV | Mains power quality | |||||
IEC | line to line | line to line | should be that of a typical | |||||
| ±2 kV | ±2 kV | commercial | or | hospital | |||
| environment. |
|
| |||||
| line to earth | line to earth |
|
| ||||
|
|
|
|
|
| |||
|
|
|
| |||||
Voltage dips, short | < 5% UT | < 5% UT | Mains power quality should | |||||
interruptions and | (> 95% dip in UT) | (> 95% dip in UT) | be that of a typical |
| ||||
voltage variations | for 0.5 cycle | for 0.5 cycle | commercial or hospital | |||||
on power supply | 40% UT | 40% UT | environment. If the user of | |||||
input lines | the | A&D unit | requires | |||||
(60% dip in UT) | (60% dip in UT) | |||||||
IEC | continued | operation during | ||||||
for 5 cycles | for 5 cycles | |||||||
| power mains | interruptions, | ||||||
| 70% UT | 70% UT | ||||||
| it is recommended that the | |||||||
| (30% dip in UT) | (30% dip in UT) | ||||||
| A&D unit be powered from | |||||||
| for 25 cycles | for 25 cycles | ||||||
| an | uninterruptible | power | |||||
| < 5% UT | < 5% UT | ||||||
| supply or a battery. |
| ||||||
| (> 95% dip in UT) | (> 95% dip in UT) |
|
|
|
|
| |
| for 5 s | for 5 s |
|
|
|
|
| |
|
|
|
| |||||
Power frequency | 3 A/m | 3 A/m | Power frequency magnetic | |||||
(50/60 Hz) |
|
| fields should be at levels | |||||
magnetic field |
|
| characteristic of a typical | |||||
IEC |
|
| location in a typical |
| ||||
|
|
| commercial or hospital | |||||
|
|
| environment. |
|
|
NOTE : UT is the AC mains voltage prior to application of the test level.
A&D Medical |