A&D UA-1020CN instruction manual Should

Models: UA-1020 UA-1020CN

1 100
Download 100 pages 1.04 Kb
Page 30
Image 30

EMC TABLE

Guidance and manufacturer’s declaration – electromagnetic immunity

The A&D unit is intended for use in the electromagnetic environment specified below. The customer or the user of the A&D unit should assure that it is used in such an environment.

Immunity

IEC 60601

Compliance

Electromagnetic

 

test

test level

level

environment – guidance

 

 

 

 

Electrostatic

± 6 kV

± 6 kV

Floors should be wood,

discharge (ESD)

contact

contact

concrete or ceramic tile.

IEC 61000-4-2

± 8 kV

± 8 kV

If floors are covered with

 

synthetic material, the

 

air

air

 

relative

humidity

should

 

 

 

 

 

 

be at least 30%.

 

 

 

 

 

Electrical fast

± 2 kV for power

± 2 kV for power

Mains power quality

transient/burst

supply lines

supply lines

should be that of a typical

IEC 61000-4-4

± 1 kV

± 1 kV

commercial

or

hospital

 

environment.

 

 

 

for input/output

for input/output

 

 

 

 

 

 

 

 

 

lines

lines

 

 

 

 

 

Surge

± 1 kV

± 1 kV

Mains power quality

IEC 61000-4-5

line to line

line to line

should be that of a typical

 

±2 kV

±2 kV

commercial

or

hospital

 

environment.

 

 

 

line to earth

line to earth

 

 

 

 

 

 

 

 

 

 

 

 

Voltage dips, short

< 5% UT

< 5% UT

Mains power quality should

interruptions and

(> 95% dip in UT)

(> 95% dip in UT)

be that of a typical

 

voltage variations

for 0.5 cycle

for 0.5 cycle

commercial or hospital

on power supply

40% UT

40% UT

environment. If the user of

input lines

the

A&D unit

requires

(60% dip in UT)

(60% dip in UT)

IEC 61000-4-11

continued

operation during

for 5 cycles

for 5 cycles

 

power mains

interruptions,

 

70% UT

70% UT

 

it is recommended that the

 

(30% dip in UT)

(30% dip in UT)

 

A&D unit be powered from

 

for 25 cycles

for 25 cycles

 

an

uninterruptible

power

 

< 5% UT

< 5% UT

 

supply or a battery.

 

 

(> 95% dip in UT)

(> 95% dip in UT)

 

 

 

 

 

 

for 5 s

for 5 s

 

 

 

 

 

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic

(50/60 Hz)

 

 

fields should be at levels

magnetic field

 

 

characteristic of a typical

IEC 61000-4-8

 

 

location in a typical

 

 

 

 

commercial or hospital

 

 

 

environment.

 

 

NOTE : UT is the AC mains voltage prior to application of the test level.

E-29

A&D Medical (Toll-Free): 1-888-726-9966

Page 30
Image 30
A&D UA-1020CN instruction manual Should