5.6. Calibration Error Code List

Exiting from a Calibration error

ESC key..... The key to return to the point where an error occurred. Retry the operation.

ESC key while pressing the ON/OFF key.

No parameters are changed, CanCel is displayed and the calibration mode is finished. Press the ON/OFF key to turn the display off.

Error Code List

If an error has occurred during the calibration mode, the following code is displayed.

Error code

Description

 

 

err 0

In multi-interval scale. The last division is set to maximum ( d-50 ).

Therefore the next division can not be entered.

 

 

Resolution exceeds 40000. (Resolution = maximum capacity/ minimum

err 1

division)

 

Reduce the maximum capacity or increase the minimum division.

err 2

Load cell output is too large or too small at zero calibration. Check the weighing

unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input

err 3

Sensitivity".

 

err 4

Measuring calibration mass, the value exceeded maximum capacity. Reduce

the calibration mass.

 

err 5

The selected calibration mass is smaller than the minimum division.

err 6

The new input sensitivity is less than 0.2 µV/division. Increase the input

sensitivity. Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity".

 

 

Placing a mass on the weighing unit, the load cell output becomes a negative

err 7

value. Check the load cell cable connections and the direction of load cell

 

mounting.

 

The load cell output exceeds the input range before the maximum capacity.

err 8

Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input

Sensitivity". Replace with a load cell designed for a smaller output. Reduce

 

 

maximum capacity.

err 9

The weight value is out of the input range at zero calibration or span

calibration. Check the weighing unit and cables.

 

err 12

The first weighing range is larger than second weighing range.

err 13

An incorrect mass was selected at the digital linearization function.

Select a mass of the following relation. Lnr 1 < Lnr 2 < Lnr 3.

 

g err

An unacceptable value was selected in the gravity acceleration function.