5.6. Calibration Error Code List
Exiting from a Calibration error
ESC key..... The key to return to the point where an error occurred. Retry the operation.
ESC key while pressing the ON/OFF key.
No parameters are changed, CanCel is displayed and the calibration mode is finished. Press the ON/OFF key to turn the display off.
Error Code List
If an error has occurred during the calibration mode, the following code is displayed.
Error code | Description | |
|
| |
err 0 | In | |
Therefore the next division can not be entered. | ||
| ||
| Resolution exceeds 40000. (Resolution = maximum capacity/ minimum | |
err 1 | division) | |
| Reduce the maximum capacity or increase the minimum division. | |
err 2 | Load cell output is too large or too small at zero calibration. Check the weighing | |
unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input | ||
err 3 | ||
Sensitivity". | ||
| ||
err 4 | Measuring calibration mass, the value exceeded maximum capacity. Reduce | |
the calibration mass. | ||
| ||
err 5 | The selected calibration mass is smaller than the minimum division. | |
err 6 | The new input sensitivity is less than 0.2 µV/division. Increase the input | |
sensitivity. Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity". | ||
| ||
| Placing a mass on the weighing unit, the load cell output becomes a negative | |
err 7 | value. Check the load cell cable connections and the direction of load cell | |
| mounting. | |
| The load cell output exceeds the input range before the maximum capacity. | |
err 8 | Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input | |
Sensitivity". Replace with a load cell designed for a smaller output. Reduce | ||
| ||
| maximum capacity. | |
err 9 | The weight value is out of the input range at zero calibration or span | |
calibration. Check the weighing unit and cables. | ||
| ||
err 12 | The first weighing range is larger than second weighing range. | |
err 13 | An incorrect mass was selected at the digital linearization function. | |
Select a mass of the following relation. Lnr 1 < Lnr 2 < Lnr 3. | ||
| ||
g err | An unacceptable value was selected in the gravity acceleration function. |
Page 21 | 5.6. Calibration Error Code List |