16.3. Command Format

Explanation of Commands

When performing a command, the received command or replay data is sent back.

When the received command can not be performed such as the "busy" state, the code " I " is sent back. In this case, insert a delay time. Electrical noise may be the cause of this error.

When receiving an undefined command (unknown command), a "?" is sent back. When there is insufficient memory to store the data of the UFC commands, an “M” is sent back.

Optional addresses can be appended to a command. The address form is "@address" and the address is specified at F-Function F06. The reply (data or error code) is also sent with the address.

Example: The command is "Display net data". The address is 23.

Sending a command

Sending a command with address

M N

Terminators

Command

@2 3 M N Terminators

Command

Address number

Unperformed command reply

I

Terminators

Unperformed command

Memory overflow (for UFC commands)

M

Terminators

Memory overflow error

Unknown command reply

@ 2 3 ?

Terminators Unknown command

Address number

Carriage return(0Dh) Line feed(0Ah)

16.3. Command Format

Page 58

AD-4405 Weighing Indicator

Page 60
Image 60
A&D AD-4405 manual Command Format, Explanation of Commands