A&D AD-4405 manual Functions, Off

Models: AD-4405

1 80
Download 80 pages 1.51 Kb
Page 25
Image 25

6.2. F-Functions

Weighing Conditions (Digital Filter, Zero trucking and Stability)

Item

Parameter

 

Description

 

0

2 d/ 1.6s

 

 

 

 

 

 

1

4 d/ 1.6s

 

 

 

 

 

 

2

8 d/ 1.6s

 

 

 

 

 

 

3

16 d/ 1.6s

 

 

 

 

 

 

4

32 d/ 1.6s

 

 

 

 

 

f00

5

64 d/ 1.6s

If weak filter is set, the response

Filter

6

128 d/ 1.6s

will be fast, but will be more

Motion / Averaging

7

2 d/ 3.2s

sensitive to external influences

time

**8

4 d/ 3.2s

such as vibration.

 

9

8 d/ 3.2s

 

 

 

 

 

 

10

16 d/ 3.2s

 

 

 

 

 

 

11

32 d/ 3.2s

 

 

 

 

 

 

12

64 d/ 3.2s

 

 

 

 

 

 

13

128 d/ 3.2s

 

 

 

 

 

 

0

OFF

This function traces the weight

 

1

0.5 d/ 1s

 

value drifting around the zero

 

2

1.0 d/ 1s

 

point slowly, displayed as zero. If

 

3

1.5 d/ 1s

 

a strong parameter is set, a very

f01

4

2.0 d/ 1s

small zero drift may be not

5

2.5 d/ 1s

Zero tracking

detected.

6

0.5 d/ 2s

 

If Cf00 1 is, the f01 0, 1, 6 or 7

 

7

1.0 d/ 2s

 

can be selected only.

 

**8

1.5 d/ 2s

 

If Cf00 1 is, the initial setting is

 

9

2.0 d/ 2s

 

f01 7.

 

10

2.5 d/ 2s

 

 

 

 

 

 

 

0

No motion detection

The function to set the condition

 

1

0.5 d/ 0.5s

of judgment whether a weight

 

2

1.0 d/ 0.5s

value is unstable or stable. The

 

3

2.0 d/ 0.5s

 

 

 

 

key are

f02

 

ZERO

key and

TARE

4

3.0 d/ 0.5s

active in the stable state. If these

STABLE mark

5

4.0 d/ 0.5s

keys need to be active in the

Motion detection

6

0.5 d/ 1s

unstable state, set to f02 0 or

condition

7

1.0 d/ 1s

select Cf04 1 or Cf04 3. In case

 

 

**8

2.0 d/ 1s

of Cf00 1 is, f02 6 or f02 7 can

 

9

3.0 d/ 1s

be selected only. (Initial setting is

 

10

4.0 d/ 1s

f02 6.)

 

0

Once after the mark

Set the number of times when

f03

 

 

the STABLE mark turned on in

1

Twice in succession

Auto Printing/

 

 

succession, until output/print out.

*2

Three times

Auto accumulation

If Cf00 1 is set, f03 2 or f03 3

 

 

3

Four times

 

can be selected.

d:division (weighing interval) of first range. s: second. *: Initial settings.

AD-4405 Weighing Indicator

Page 23

6.2. F-Functions

Page 25
Image 25
A&D AD-4405 manual Functions, Off