8.1.2.Display and Operation

Action of Accumulating Data

 

 

 

When accumulating data, the display blinks once.

 

 

 

 

 

 

 

 

 

If the accumulated data is stored, the M+ mark is displayed.

Caution

 

 

This function can not accumulate data with a different unit. Specify a unit

 

 

 

 

 

 

 

 

 

before use.

Display of Accumulated Data

When setting to Cf08 1 (Effective accumulation function) and pressing the TOTAL

key, total is displayed and the total data is displayed with the M+ mark blinking. Pressing the TOTAL key alternatively displays the accumulated data and the accumulation count. Pressing the ESC key, the weight data is displayed.

The total data can be output. Refer to "Output of Accumulation Data".

Undoing the Accumulation Data

The last weight data can be deduced from the accumulated data unless new data has been accumulated.

Step 1

Press the

TOTAL

key to display

total

and the accumulated data.

Step 2

Press and hold the +/- key for more than 3 seconds. The display blinks once and

 

the data accumulated before accumulating the last weight data is displayed.

Caution

 

 

External input can not be used.

 

 

 

 

Clearing the Accumulation Data

 

 

 

 

Step 1

Press the

TOTAL

key to display

total

and the accumulated data.

 

 

 

 

Step 2

Press and hold the

ZERO

key more than 3 seconds. The display blinks once and

 

the accumulated data is cleared.

Caution

 

 

External input can not be used.

 

 

 

 

Initializing the Data Number and Clearing the Accumulation Data at the same time

When the data number is included with the data of the serial data output or the printer, initializing the data number and clearing the accumulated data can be done at same time.

If using both serial output and printer data number, all data numbers are initialized by this operation.

Step 1

Press

TOTAL

key to display

total

and the accumulated data.

 

 

 

 

 

 

Step 2

Press and hold the

ZERO

and +/- keys at the same time for more than 3 seconds.

 

The display blinks once and the accumulated data is cleared. And the data

 

number is initialized (1).

Caution

 

 

External input can not be used.

 

 

 

 