A&D AD-4405 manual Data Number, Interval Data Output / Print

Models: AD-4405

1 80
Download 80 pages 1.51 Kb
Page 56
Image 56

Description of "Command Mode" (For Serial Interface)

Function

f40 5

Operation

The serial data can be output by the serial interface command only. Data

 

conflicts or accidental data output by mis-operation can be avoided when

 

using multiple instruments with the RS-422/485 interface.

Description of "Delay for continual data"

Function

f38, f42

Operation

This function can be used in the "Auto print mode" and "Manual print mode".

 

When using a non-buffered printer, set to f42 3 and f43 1 (or f38 1).

15.2. Data Number

The data number increments automatically with each serial data output or print out respectively from 1 to 99999 (the next of 99999 is 1). The starting number can be set using the keys.

Setting the Serial Data Number

Press the CODE(1) key while pressing the PRINT key at the weight display, the display changes to data r then r xxx. “xxx” is the next data number. Use the 0 - 9 keys to enter the number, then press the ENTER key to store the number and return to the weight display.

Setting the Print Data Number

Press the SAMPLE(2) key while pressing the PRINT key at the weight display, the display changes to data p then p xxx. “xxx” is the next data number. Use the 0 - 9 keys to enter the number, then press the ENTER key to store the number and return to the weight display.

Note 1. At the number setting, the following keys are available.

+/- key

Sets the number to 1.

 

 

 

Sets the number to 0.

ZERO

key

 

 

Returns to the weight display without changing the number.

ESC

key

2. When setting the number to 0, it is entered as 1 automatically.

15.3. Interval Data Output / Print

The AD-4405 can output and/or print data with a constant time interval using the internal calendar/clock function. The time intervals for serial data output and for the printer can be set individually from 1 second up to 23 hours and 59 minutes.

Caution The interval time must be set to allow enough time to finish output or printing out.

15. Common Printer and Data Output

Page 54

AD-4405 Weighing Indicator

Page 56
Image 56
A&D AD-4405 manual Data Number, Interval Data Output / Print