5.2.3.Zero Calibration

Procedure

Step 8 Check the

Cal 0

display.

Zero calibration

Select a zero calibration method to adjust the zero point

Weighing input

The adjustment method

 

with nothing on the

To step 9

(Normal way)

weighing unit.

 

 

 

 

The numerical way to

 

Digital input

enter a load cell output

To step 10

 

voltage.

 

Weighing Input

Step 9 Place nothing on the weighing unit. Press the ENTER key after the STABLE mark has turned ON. The new zero point parameter will be stored. Proceed to step11. ESC key .......... The key not to change the zero point

data and proceed to the next step.

Caution: Do not press the ENTER key while the STABLE mark is off (detecting motion). Arrange the condition using the F00 filter function.

Digital Input

Step10 Pressing the +/- key, a stored input voltage parameter of the zero point is displayed in the unit of mV/V. Adjust the input voltage using the 0 - 9 keys.

ENTER....... The key to store the zero point parameter

Place nothing on the weighing unit (No load)

Turn on STABLE mark

PRINT

ENTER

Stored input voltage

NET/GROSS

+/-

mV/V

and proceed to the next step.

ESC key..... The key to proceed to the next step without

changing the parameter.

0 9

mV/V

PRINT

New input voltage

ENTER

 

To Span Calibration

5.2.4.Span Calibration

Step11 Check the capacity display after Cal f is displayed for 2 seconds.

Select a span calibration method to adjust the capacity.

Weighing a mass less than the

The method to weigh a mass less

To step 12

maximum capacity

than the maximum capacity.

 

Weighing a maximum capacity

The method to weigh a mass

To step 14

mass

equivalent to the maximum capacity.

 

Digital input

The numerical way to enter a load

To step 16

cell output voltage.

 

 

AD-4405 Weighing Indicator

Page 15

5.2. Calibration Procedure