A&D AD-4405 Adjustment of the Load Cell Output, Verifying Load Cell Output and Input Sensitivity

Models: AD-4405

1 80
Download 80 pages 1.51 Kb
Page 8
Image 8
A change in input voltage sensitivity is required to cause a one division change of the display. Select as large an input sensitivity voltage as possible so that the weighing interval becomes stable.
Consider the leverage if a lever is used.

3.1.3.Adjustment of the Load Cell Output

Caution

Use a metal film resistor in the range of 50kohm to 500kohm with a good temperature coefficient, when adding a resistor to adjust a load cell output. Use as large of a resistance value as possible in the range in which the zero adjustment is possible. Solder this resistor at a point near the load cell or the indicator.

In Case of Reducing the Output Voltage

When the zero output is too large, add a resistor between EXC+ and SIG-.

In Case of Adding an Offset Voltage to the Output When the zero output is too small, add a resistor between EXC+ and SIG+.

3.1.4.Verifying Load Cell Output and Input Sensitivity

The input sensitivity of the indicator is 0.25mV /division or more. Adapt to the following inequality, when you design a weighing instrument using the indicator and load cell(s).

Caution

 

 

 

 

 

 

E * B * D

 

 

Weighing instrument

 

0.25 £

 

 

 

 

 

using one load cell.

A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weighing instrument

 

0.25 £

E * B * D

 

 

using multi-load cell

A * N

 

 

 

 

 

 

 

 

 

 

 

 

 

A:Rated capacity of load cell [kg]

B:Rated output [mV/V]

D:Weighing interval [kg]

E:Excitation voltage [mV]

N:Number of load cells

Verification Example

 

Design:

 

 

 

 

 

 

 

 

Load cell

N=1

 

5000

* 3 * 0.05

 

 

 

Rated capacity

A=750 [kg]

 

= 1 ³ 0.25

. Therefore,

 

 

 

 

 

 

 

750

 

Rated output

B=3 [mV/V]

 

 

 

 

 

 

 

regard the instrument as a good

 

Excitation voltage

E=5000 [mV]

 

 

 

design.

 

 

Weighing interval

D=0.05 [kg]

 

 

 

 

 

 

 

 

 

 

Weighing capacity

300 [kg]

 

 

 

 

 

 