19. 4-20mA Analog Output (OP-07)

The OP-07 analog output option is for sending the weight data to an analog input unit. The output is a 4mA to 20mA current output proportional to the display reading.

The output data is updated in synchronization with the display update.

Specifications

Output current

4mA to 20mA *

 

Non-linearity

 

Less than +/- 0.1% fs

Load resistance

0 to 510 W

 

Temperature

ZERO

Less than +/- 0.02% fs/°C

Resolution

Approx. 1/10000

 

coefficient

SPAN

Less than +/- 0.02% fs/°C

Output terminal

Connector terminal

No.1 : + No.2 : - No.3 : FG(Earth)

Accessory

Connector terminal (1) TM-MSTB03STF

 

 

AC adapter (1) Note: Please confirm that the AC adapter type is correct

 

 

 

for your local voltage and receptacle type.

*When set to non-weight display (Calibration, F-settings etc.), output current is 4mA.

The output current is not adjustable.

+ - ￿￿ ￿￿ ￿

 

&%

 

 

￿￿

￿￿

()

 

 

 

 

 

 

 

 

 

DC 7V - 10V

Analog output terminal

 

 

 

 

 

AC adapter Jack

 

 

 

 

 

Settings

Set f30 1 of the F-Function, and set f31, f32 and f33.

 

Item

Parameter

Remarks

 

 

 

 

 

 

 

 

 

 

 

Data

f30

* 0

No output

 

Initial setting

 

 

output

Data output

1

Analog output

 

Must be set to 1

 

 

 

2

Serial in/out 1

 

 

 

 

 

 

3

Serial in/out 2

 

 

 

 

Analog

f31

* 0

Displayed value

 

Initial setting

 

 

output

Output data

1

GROSS weight

 

 

 

 

 

 

2

NET weight

 

 

 

 

 

f32

-999999 to 999999

 

Decimal point

is

set

 

Weight value at 4mA

(Initial setting is 0)

 

at “CALSEt”

 

 

 

f33

-999999 to 999999

 

Decimal point

is

set

 

Weight value at 20mA

(Initial setting is 10000)

 

at “CALSEt”

 

 

Settings of f32 and f33

When entering the f32 or f33 settings (press the

ENTER key when “32” or “33” is blinking), the setting value is displayed. Set the value using the 0 - 9 key. By pressing the +/￿￿ key, the polarity of the value can be alternated. Press

the ENTER key to store the setting value into memory. After this the display returns to selection of the Function number.

AD-4405 Weighing Indicator

Page 69

Output current

20mA

 

b

a

4mA

 

f32 a

f33 a

f33 b

f32 b Weight