8. Accumulation

This function accumulates weighing data and stores the total data and the accumulation count. The data is stored in non-volatile memory, and is not lost even if the power is turned off.

8.1.1.Preparation and Specifications

Set the following parameters to use the accumulation function.

Select Cf08 1 for the CF-function so that the accumulation function becomes effective. Specify the method of accumulation and data at f20 of the F-functions.

Specify the inhibit region for accumulation at f21 of the F-functions.

Selection of Accumulation Mode, F20 of the F-functions

There are two methods of accumulation; manual accumulation using the M+ key and automatic accumulation.

The accumulated data can select "positive data only" or "both polarity data". The accumulated data can select "result of comparison is OK only" or "all result".

Accumulation Condition, F21 of the F-functions

In the case of manual accumulation mode, press the M+ key to accumulate weight data when the STABLE mark is displayed.

Data can be accumulated after the weight data enters the "inhibit region for output". When connecting the power cord and turning the display on, the accumulation mode

takes the same action.

Inhibit region for accumulation

f21

 

Description

Add data anytime

f21

0

Stable data can be used anytime

Above

 

5 d

f21

1

Initial setting

 

Above

 

10 d

f21

2

 

 

 

Above

 

20 d

f21

3

 

 

 

Above

 

50 d

f21

4

 

 

 

Caution

Do not set f21 0 for the automatic accumulation mode

If setting f21 0, it may add the same data two times or more.

Limitation of Accumulation Count and Total

The limitation of accumulation count is 999999.

The limitation of total is 999999 ignoring the decimal point.

If exceeding these limitations, the data is not accumulated.

Example: Setting the decimal point to "0.0", the limitation is "99999.9".