External Input

 

Item

 

Parameter

 

 

 

Description

 

 

 

** 0

Not used (No function)

 

 

 

 

 

 

1

ZERO key

 

 

 

 

 

 

 

2

TARE key

 

 

 

 

 

 

 

3

NET/GROSS key

 

 

 

 

 

 

4

ON/OFF key

 

 

 

 

 

 

5

PRINT key, ENTER key

 

 

 

 

 

 

6

FEED key

 

 

 

 

 

 

 

7

Serial data output (Format 1)

 

 

 

 

 

8

Serial data output (Format 2)

 

 

f13

 

 

9

Accumulation (M+)

 

 

 

EXT1

 

 

10

Start batching

 

 

 

Function selection of

 

11

Stop batching

 

 

 

external input

 

12

“Over” signal, Gross over and display data out when ON

 

 

 

13

NET weight display when shorting the terminal.

 

 

 

14

Accumulated data display when shorting the terminal.

 

 

 

15

Execute comparison when shorting the terminal.

 

 

 

16

Display by the second unit when shorting the terminal.

 

 

 

17

All keys are enabled when shorting the terminal.

 

 

 

 

(Ignores F12 setting temporary)

 

 

 

18

Inhibit memory code reading when shorting the terminal.

 

 

 

19

Start averaging to hold

 

 

 

 

 

 

20

Release holding data

 

 

 

f14

 

 

0 to 17, 19 and 20

Same as f13

 

 

 

EXT2

 

 

18

Code Memory (BCD 1)

 

 

 

f15

 

 

0 to 17, 19 and 20

Same as f13

 

 

 

EXT3

 

 

18

Cord Memory (BCD 2)

 

 

 

*:

Initial settings.

 

 

 

 

 

 

 

 

Accumulation

 

 

 

 

 

 

 

 

 

Item

 

Parameter

 

 

 

Description

 

 

 

10 11

Acc. no (0) / Exec (1)

Left: Sub item, select using the

 

 

 

 

 

 

 

 

 

ZERO

key

f20

[Type2]

 

20 21

Manual (0) / Auto (1)

 

 

Right: Setting, select using the

Accumulation Mode

 

30 31

+ Only (0) / +/- (1)

 

 

0 , 1 keys

 

 

 

 

 

 

 

 

 

 

 

 

40 41

OK only (0) /All data (1)

 

 

 

 

Initial settings: 10, 21, 31, 41

 

 

 

0

Add data anytime

The selection of the inhibit region

f21

 

 

** 1

Above

 

5 d

 

 

 

for accumulation.

 

 

 

Inhibit region for

 

2

Above

 

10 d

 

 

Do not set f21 0 when Auto

accumulation

 

 

 

 

 

 

 

3

Above

 

20 d

 

 

accumulation (f20 21).

 

 

 

 

 

 

 

 

 

 

 

4

Above

 

50 d

 

 

 

 

 

 

 

 

 

 

 

 

 

 

d: division (weighing interval) of first range.

*:

Initial settings.

AD-4405 Weighing Indicator

Page 25

6.2. F-Functions

Page 27
Image 27
A&D AD-4405 manual EXT1, EXT2, EXT3