ESC key ..... The key to store the parameters temporarily. Proceed to the Cal 0

display.

Press and hold the ON/OFF key and press the ESC key

No parameters are changed, CanCel is displayed and the calibration mode is finished.

Step18 Press the ON/OFF key to turn the display off.

5.3. Weighing Range Function

The weighing range function can select "single range" or "dual range". Specify each weighing interval (division) for the multi-interval instrument. Each weighing interval is displayed according to a net value or gross value.

Caution

 

 

 

 

When single range is used, performing this function is not required.

 

 

 

 

 

 

 

 

Example 1

 

 

 

 

The gross display.

 

 

 

 

 

 

 

Specified parameters:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

First range

 

 

Range = 50.00kg, division 0.02kg

 

 

 

 

 

 

 

Second range

 

 

Range = 100.00kg (maximum capacity), division 0.1kg

Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0kg to 50kg :

 

 

The first range, division 0.02kg.

 

 

 

 

 

 

 

 

 

 

 

 

 

50kg to 100kg :

 

 

The second range, division 0.1kg.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Division = 0.02kg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Division = 0.1kg

 

0kg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50kg

 

 

 

100kg Gross

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weighing value

Example 2

 

 

 

 

The net display using a 40kg tare value.

 

 

 

 

 

 

 

Specified parameters:

 

 

The same parameters as example 1.

 

Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-40kg to 50kg :

 

 

The first range, division 0.02kg.

 

 

 

 

 

 

 

 

 

 

 

 

 

50kg to 60kg :

 

 

The second range, division 0.1kg.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Division = 0.02kg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Division=0.1kg

 

 

-40kg

 

 

0kg

 

50kg 60kg

 

Net

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gross

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Tare value

 

90kg 100kg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

40kg

Weighing value

 

AD-4405 Weighing Indicator

Page 17