13. Calendar / Clock

The AD-4405 has a built-in calendar/clock backed up by a lithium battery. The calendar/clock continues to operate with AC power turned off. Leap years are automatically corrected. The time and date can be added to the data output and optional built-in printer output.

 

 

 

13.1.

Time and Date Functions

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Item

 

Parameter

Description

 

 

f50

 

*0

12hours (00-11 AM/PM)

Display, serial data output

 

 

Time display

 

1

24hours (00-23)

and print output

 

 

f51

 

*0

YY/MM/DD

Display, serial data output

 

 

 

1

MM/DD/YY

 

 

Date format

 

and print output

 

 

 

2

DD/MM/YY

 

 

 

 

 

 

 

 

 

f52

 

00 to

(Display/output year) –

Last 2 digits

 

 

Difference between

 

 

 

 

99

(Christian year)

Initial setting is 0.

 

 

Christian year

 

 

 

 

 

 

 

 

 

 

 

 

 

*0

2digits

 

 

 

f53

 

1

4digits (Numeric)

First 2digits specified at

 

 

 

f54.

 

 

Print/output of year

 

 

 

 

 

 

2

4digits (alphabets and year)

Specify alphabet and

 

 

 

 

 

 

 

 

 

 

 

 

format at f54.

 

 

 

 

 

 

 

 

 

 

f54

 

00 to

2digits number (f53 1) or

 

 

 

First 2ditigts of

 

Effective f53 1 or 2

 

 

 

99

alphabet, blank or dot

 

 

year

 

 

 

 

 

 

 

 

 

f52 : The display or output year is calculated using the following formula. Display/output year = (Christian era) + (f52)

Example 1: Christian era = (20)05, f52 10 then display/output year = (20)15

Example 2: Christian era = (20)05, f52 99 then display/output year = (20)04 Keep the internal year in Christian era format because the leap year correction is made based on the Christian era.

f53 and f54 : 2 characters can be added preceding the internal year to the output of the year. Additional characters are set at f54. A 4 digit year can not be displayed.

f53 0 Add no character, output year is 2 characters.

f53 1 Add 2 digit numeric characters. The 2 digits number does not increment automatically if the internal year changes from 99 to 00.

Example: Setting f54 20, internal year (after f52 calculate) = 05, output year = 2005