15. Common Printer and Data Output Items

There are some similar functions and operations for the internal printer and the serial data output (RS-232C, RS-422/485 and the current loop).

Data output / print output mode

Data number (not available the current loop)

Interval data output or print out

 

 

 

15.1.

Data Output / Print Mode

 

 

 

 

 

 

 

 

 

Description of "Stream Mode" (For data output)

Function

f37 0, f40 0

 

 

 

 

Operation

The data is output at each sampling (when refreshing the display).

 

 

 

 

 

Use this mode to output data to an external display (The data may not be

 

 

 

 

 

output due to timing of the baud rate and internal sampling rate). If the data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

is to printed by pressing the

PRINT

key on the printer, use the stream mode.

 

 

 

 

 

The serial interface command can not be used at f40 0.

Description of "Manual Print"

 

 

 

 

Function

f37 1, f40 1, f60 1

 

 

 

 

 

 

 

 

 

 

Operation

When pressing the

PRINT

key or the external input is set to output or print,

 

 

 

 

 

the stable weighing data is output and/or print just once.

 

 

 

 

 

The serial interface command can be used at f40= 1.

Description of "Auto Print Mode"

 

 

 

 

 

 

 

 

 

Object

f37 2 or 3, f40 2 or 3, f60 2 or 3,

Operation

When the weight data varies from the "inhibit region for output" to the

 

 

 

 

 

"permission region of output", the stable data is output and/or print just once.

 

 

 

 

 

If you use this mode, set f02 1 ~ 10, but not 0.

 

 

 

 

 

The serial interface command can be used at f40 2 or 3.

 

 

 

 

 

For weighing (and removing) each object and printing the data.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In case of f37 2, f40 2, f60 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

"Inhibit region for output" * +5d.

+5d < "permission region of output".

 

 

 

 

 

In case of f37 3, f40 3, f60 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-5d * "inhibit region for output" * +5d.

 

 

 

 

 

"Permission region of output" < -5d, +5d < "permission region of output".

 

 

 

 

 

 

 

 

 

d: division (weighing interval) of the first range.

Description of "Accumulation Data Output/Print"

Function

f37 4, f40 4, f60 4

 

 

 

 

Operation

When accumulating by manual or automatic, data is output and/or print

 

 

 

 

 

automatically with the accumulated data.

 

 

 

 

 

The serial interface command can be used at f40 4.

AD-4405 Weighing Indicator

Page 53

15. Common Printer and Data Output

Page 55
Image 55
A&D AD-4405 manual Common Printer and Data Output Items, Data Output / Print Mode