Weighing a Mass less than the Maximum Capacity

Step12

Set a mass value using the 0 - 9 keys.

Step13

Place a mass equivalent to the displayed value

 

on the weighing unit. Proceed to step 15.

Weighing a Maximum Capacity Mass

Step14 Place a mass equivalent to the maximum capacity on the weighing unit.

Step15 Press the

ENTER

key after the

STABLE

mark turns on. Proceed to step17.

 

 

 

..........

The key to proceed to step17

 

ESC

key

without changing the span parameters.

Caution: Do not press the ENTER key while the STABLE mark is off (detecting motion). Arrange the condition using the F00 filter function.

Digital Input

Step16 Pressing the +/- key, a stored input voltage parameter of the span is displayed in the unit of mV/V. Adjust the input voltage using

the 0 - 9 keys. (It is possible to store a greater value than the capacity.)

ENTER key...... The key to store the span

parameters and proceed to step 17.

ESC key .......... The key to proceed to step 17

without changing the span parameters.

Span calibration

Maximum capacity

To step 14, step 16

 

 

To step 12

 

0

9

New mass value

 

Mass of

 

maximum capacity

 

 

 

step 13

 

To step 14

step 14

To step 16

Place the displayed

mass on the

 

 

weighing unit

 

Turn on STABLE mark

 

 

PRINT

 

 

ENTER

NET/GROSS

Stored input voltage

 

+/-

 

mV/V

0

9

mV/V

PRINT

New input voltage

ENTER

 

5.2.5.Exiting the Calibration Mode

Step17 Check the Calend display.

Use the following keys.

CAL key ..... To store the parameters and display

Caloff. Proceed to the next step. Caloff could be displayed by pressing the +/- key while pressing the ON/OFF key instead of CAL key.

NET/GROSS

or

+/-

PRINT

ENTER

Turn display off

PRINT

ENTER

5.2. Calibration Procedure

Page 16

AD-4405 Weighing Indicator

Page 18
Image 18
A&D AD-4405 manual Exiting the Calibration Mode, Stable