Specifications

Model

UA-851THW/UA-851TMW/UA-851THX

 

 

Type

Oscillometric

 

 

Display

Digital, 20-mm character height

 

Pressure/pulse displayed simultaneously

 

 

Memory

51 readings

 

 

Measurement range

Pressure: 20 mmHg to 280 mmHg

 

Pulse: 40 pulses to 180 pulses

 

 

Accuracy

Pressure: ±3 mmHg or ±2%, whichever

 

is greater

 

Pulse: ±5%

 

 

Pressurization

Automatic, using micropump

 

 

Depressurization

Active exhaust valve

 

 

Power source

4 type “AA” (1.5 volt) alkaline

 

batteries (included) or 120 V

 

AC Adapter (optional)

 

 

Battery life

Approximately 4 months with

 

one daily measurement

 

 

Recommended operating

50ºF to 104ºF (10ºC to 40ºC)

environment

Less than 85% relative humidity

 

 

Recommended storage environment

14ºF to 140ºF (-10ºC to 60ºC)

 

Less than 95% relative humidity

 

 

Dimensions

Length: 4.7”(120 mm)

 

Width: 5.9”(150 mm)

 

Height: 6.1” (156 mm)

 

 

Weight

17.6 oz. (500 g) without batteries

 

 

Blood pressure measurements determined by the UA-851THW are equivalent to those obtained by a trained observer using the cuff/ stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.

NOTE: Specifications are subject to change without prior notice or any obligation on the part of the manufacturer.

Page E-17 A&D Medical (Toll-Free): 1-888-726-9966

Page 20
Image 20
A&D UA_851THX, UA851TMW user manual Specifications, UA-851THW/UA-851TMW/UA-851THX