Agilent Technologies 6633B CV/CC Status Annunciators Troubleshooting, Bias and Rail Voltages

Models: 6632B 6633B 66332A 6634B

1 82
Download 82 pages 25.5 Kb
Page 38
Image 38

3 - Troubleshooting

CV/CC Status Annunciators Troubleshooting

The CV/CC annunciators are particularly helpful when troubleshooting a unit with no output. If the unit has no output voltage or current and one of the annunciators is on then the problem is in the control circuit associated with that annunciator. An example of how this might be useful would be in a case where the voltage and current are programmed to some positive value, there is no output voltage and the CV annunciator is on. This indicates that the problem is probably in the Voltage Amplifier circuit. If the CC annunciator were on then the problem would likely be in the Current Amplifier. If UNR is indicated then neither the voltage nor the current circuits are in control and the problem would be in circuits after the gating diodes such as the driver or output regulator stages.

When troubleshooting the CV/CC status annunciators or the status readback circuits, first measure the voltage drop across the gating diodes; A1 D328 (CV) and D325 (CC). A conducting diode indicates an active (ON) control circuit. This forward drop is applied to the input of the associated status comparator (U306A and D respectively) and drives the output (CV_DETECT* or CC_DETECT*) low. The low signal indicates an active status which is sent to the A2 board microprocessor. The front panel CV annunciator indicates when the CV mode is active (CV_DETECT* is low). The front panel CC annunciator indicates when the CC mode is active (CC_DETECT* is low). The UNREGULATED (UNR) annunciator comes on when neither the CV nor CC is active.

Bias and Rail Voltages

Before troubleshooting any circuit check the bias and/or rail voltages to make sure that they are not the cause. Table 3-3 lists the bias and rail voltage test points for the A1 Main Control , A2 Interface, and the A3 Front Panel/Display boards. Unless otherwise noted, all voltages are measured with respect to secondary common (R473-3) with no load on the supply.

Table 3-3. Bias and Reference Voltages

Bias

Test Point

Common

Measurement

 

(See Figure 6-1)

 

 

 

+Rail1 (Agilent

A1 TP 310

- Output

+38V

10% (800mV P/P)

6632B/66332B)

 

 

 

 

+Rail1 (Agilent 6633B)

A1 TP 310

- Output

+73V

10% (2.5V P/P)

+Rail1 (Agilent 6634B)

A1 TP 310

- Output

+130V 10% (2.2V P/P)

-Rail1(Agilent

A1 TP 311

- Output

-9.8V 10% (400mV P/P)

6632B/66332B)

 

 

 

 

-Rail1(Agilent 6633B)

A1 TP 311

- Output

-10.2V 10% (300mV P/P)

-Rail1(Agilent 6634B)

A1 TP 311

- Output

-10.5V 10% (300mV P/P)

+5V secondary

A1 R317

Secondary Common

+5V 4%

+12V secondary

A1 D470 cathode

Secondary Common

+12V

5%

+15V secondary

A1 R318

Secondary Common

+15V

4%

-12V secondary

A1 D471 anode

Secondary Common

-12V 5%

-15V secondary

A1 R315

Secondary Common

-15V 4%

V_Ref

A1 R475

Secondary Common

+2.5V 6%

+5V Interface2

E306 (red wire)

E 306 (black wire)

+5V 3%

1Measured with respect to - Output at nominal ac input line voltage

2Measured with reference to Interface Ground (E306 black wire)

38

Page 38
Image 38
Agilent Technologies 6633B, 6634B, 66332A, 6632B CV/CC Status Annunciators Troubleshooting, Bias and Rail Voltages