2-11
Using the Agilent 83434A
BER Performance Verification
2Set the Agilent 71612B as follows:
CLOCK OUTPUT
SIG GEN FREQ. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9953.28 MHZ
SIG GEN FREQ (for 83434 option 106) . . . . . . . . . . . . . . . . . . . . . . . . 10664.23 MHZ
MENU
DATA OUTPUT
Ext AC COUPLED
DATA AMPLITUDE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1.0 V
PATTERN
PRBS
2 31–1
3Disable WAVELENGTH ADJUST on the 83433A.
4Turn on the 83433A laser.
5Adjust the optical attenuator for a maximum of 0 dBm and a minimum of
16 dbm at the output of the attenuator.
6Set the Agilent 71612B as follows:
MENU
INPUT & EYE
0/1 THR CENTER
CLK/DATA ALIGN
MENU
GATING
RUN GATING
7Verify that the 71612B reports zero errors.