4-4
Specifications and Regulatory Information
Agilent83434A Specifications and Characteristics
Recovered Clock Output
Amplitudeh
Frequencyk
Frequency (opt. 106)
Duty cycle
Clock to data alignmentl
3 dB bandwidthk
Jitter generationg,m
Return lossn
Impedance
0.5 to 1. 5 V pk-pk
9953.26 to 9953.30 MHz; 9953.28 nominal
10664.03 to 10664.43 MHz; 10664.23 nominal
45/55% maximum; 50/50% nominal
±25.12 ps maximum
8 to 12 MHz; 10 MHz nominal
2 ps rms maximum
12 dB minimum
50
GENERAL SPECIFICATIONS
Temperature Range
Operating 0°C to +45°C
Storage 40°C to +70°C
EMI Compatibility Conducted and radiated emissions meet the
requirements of CISPR Publication 11, Class A and
immunity in compliance with IEC 61326-1
Power Requirements 100/120/220/240 V (±10%), 47 to 63 Hz
Weight (characteristic) 3.4 kg (7.6 lb)
Dimensions 102mm (4 in) height, 216 mm (8.5 in) width, 444
mm (17.5 in) depth (Agilent System II, half-width
case)
FRONT-PANEL INPUT / OUTPUT
Optical Input Connector Diamond HMS-10/HPo
Data Output Connector APC-3 .5 mal e
Recovered Clock Output Connector APC-3.5 male
a. Better than 1x10-10 BER when tested with the Agilent 71612B using either of two patterns: a) 2311 PRBS,
or b)100 consecutive ones or zeros on a 2311 PRBS.
b. Source extinction ratio 8.2 dB measured within ±10% of eye center
c. Applies over the t emperature range 25°C ±10°C.
d. Tested with FC/PC adap ter.
e. 1310 or 1550 nm.
f. Non-inverting, non-retimed linear output with AGC stabilization.
g. For PRBS up to and including 2311.
h. AC coupled
i. Measured with a swept network analyzer at 8 dBm optical input with fixed AGC control in Tx and Rx.
j. 0.01 to 10,700 MHz.
k. Type A resonator based clock recovery.