3-16
Reference
Fiber-Optic Connectors
To clean a non-lensed connector
CAUTION Do not use any type of foam swab to clean optical fiber ends. Foa m swabs can
leave filmy deposits on fiber ends that can degrade performance.
1Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotto n fibers remain on the fiber end
after cleaning.
2Clean the ferrules and other parts of the connector while avoiding the end of
the fiber.
3Apply isopropyl alcohol to a new clean lint-f ree cotton swab or lens paper.
4Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caug ht in the
swab and become a gouging element.
5Immediately dry the fiber end with a clean, dry, lint-free cotton swab or l en s
paper.
6Blow across the connector end face from a distance of 6 to 8 inches using
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the
fiber end face.
Nitrogen gas or compressed dust remover can also be used.
Table3-5. Dust Caps Provided with Lightwave Inst rume nts
Item AgilentPart Number
Laser shutter cap 08145-64521
FC/PC dust cap 08154-44102
Biconic dust cap 08154-44105
DIN dust cap 5040-9364
HMS10/dust cap 5040-9361
ST dust cap 5040-9366