Step 4. Install M-Modules onto the Carrier

1.For internally-mounted M-Modules (slots M4 and M5) see Step 5, next page.

2.Install RFI Dress Panel (p.n. E2273-04301) on each of the front mounted M-Modules. Dress panels act like a gasket and minimize RFI. Make certain the M-Module-to-Carrier connectors line up properly. Make certain the M-Module standoffs are aligned over the screw mounting holes.

3.Secure each M-Module to the Carrier (from the back side) with four M3x5 mounting screws

(p.n. 0515-0372, with captive lock washer) supplied with M-Module. Open handles on Carrier to install screws for slots M0 and M3 M-Modules.

4.Install blank Dress Panel (p.n. E2251-04301) over each unused front panel slot to minimize RFI.

2. Align M-Module on Carrier.

3. Secure M-Module with four M3x6 screws. 0515-0430

Do not remove these screws if removing M-Module.

1. Install E2273-04301 Dress Panel on M-Module.

4.Install E2251-04301 Blank Dress Panels on Empty Slots.

If necessary, slightly crimp Dress Panel with pliers so it stays attached to Carrier front panel.

Caution Static Electricity. Use anti-static techniques when installing/removing M-Module.

Use care when installing M-Modules onto the carrier connectors. The Carrier and M-Module connectors ARE NOT keyed; it is possible to misalign an M-Module and connect it to the wrong connectors. APPLYING POWER TO A MISALIGNED M-MODULE WILL DAMAGE OR DESTROY THE M-MODULE AND THE CARRIER.