E2272A Dual 8-to-1/Single 16-to-1 Relay Multiplexer Wiring Information

MAXIMUM VOLTAGE/CURRENT: The maximum voltage that may be applied to any connector on the E2270A, E2271A, or E2272A is 200 VDC, 125 VAC rms, or 175 VAC peak. These limits apply only if the product is installed in a humidity-controlled (<60% RH) environment where airborne contaminants and transients are controlled, and there is NOT a relay connection made to power mains. If these conditions CANNOT be maintained, then the maximum voltage is 60 VDC, 48 VAC rms or 68 VAC peak.

The maximum current (non-inductive) that may be applied to the E2270A, E2271A, or E2272A is:

Per Switch: 2 ADC, 2 AAC peak

Per M-Module: 8ADC, 8AAC peak

User Connector Pins

User Connector Pins

Channel 0

 

Channel 8

 

1

CH0_HI

25

CH8_HI

2

CH0_LO

9

CH8_LO

Channel 1

 

Channel 9

 

18

CH1_HI

26

CH9_HI

33

CH1_LO

10

CH9_LO

Channel 2

 

Channel 10

 

17

CH2_HI

27

CH10_HI

32

CH2_LO

11

CH10_LO

Channel 3

 

Channel 11

 

16

CH3_HI

28

CH11_HI

31

CH3_LO

12

CH11_LO

Channel 4

 

Channel 12

 

22

CH4_HI

13

CH12_HI

6

CH4_LO

29

CH12_LO

Channel 5

 

Channel 13

 

21

CH5_HI

15

CH13_HI

5

CH5_LO

14

CH13_LO

Channel 6

 

Channel 14

 

20

CH6_HI

30

CH14_HI

4

CH6_LO

44

CH14_LO

Channel 7

 

Channel 15

 

19

CH7_HI

43

CH15_HI

3

CH7_LO

42

CH15_LO

34 - 41

8

MUXB_HI_COM

 

 

 

Single/Dual Mux. Jumpers

24

MUXB_LO_COM

 

 

 

 

 

23

MUXA_HI_COM

 

 

7

MUXA_LO_COM

30