Cleaning

OPTICAL CONNECTOR

Cleaning

The glass power meter interface does not make contact with the inserted connector: there is a slight air gap. Therefore it should not wear out, and only needs occasional cleaning.

To clean the glass power meter interface: remove the interchangeable adaptor to access the glass interface, then use a sticky material such as ‘ Blu tac ‘ to dab the glass and remove any dirt. A soft brush, alcohol, or air can are also appropriate. Do not use anything hard that could scratch the glass.

In this manual, refer to the sections on Cleaning Information and Procedures for further information.

Bare fiber adaptors

Bare fiber adaptors used with this power meter must achieve

+

fiber eccentricity of _ 100 microns and end tolerance of

+

_ 0.3 mm, relative to the ferrule end. Be careful not to scratch the detector interface when using adaptors, since permanent

meter inaccuracy may result. Preferred bare fiber adaptors consist of a connector with fiber retention device.

If you work regularly with bare fibers, to avoid potential damage to the meter, it is generally preferable to use an alternative arrangement such as a multimode pigtail with a v-groove or mechanical splice.